Sensorless Physiologic Control, Suction Prevention, and Flow Balancing Algorithm for Rotary Biventricular Assist Devices
- 13 December 2017
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Control Systems Technology
- Vol. 27 (2) , 717-729
Abstract
Objective: Rotary biventricular assist devices (BiVAD) are mechanical pumps that are implanted in the left and right ventricles of biventricular failure patients to pump blood and provide mechanical circulatory support. The objective of this paper was to develop and test a novel sensorless control algorithm that simultaneously satisfies the objectives of providing physiologic control (BiVAD flows meet cardiac demand), preventing ventricular suction, and providing balanced left-right (systemic and pulmonary) flows without the use of implantable flow or pressure sensors in the nonlinear, time varying, and discontinuous circulatory system. Methods: The control algorithm consists of two gain-scheduled proportional-integral controllers for left and right ventricular assist devices and only requires intrinsic pump parameters (speed and power) to maintain differential pump speeds (Δ RPML and Δ RPMR) above user-defined thresholds to prevent ventricular suction, and average reference pressure heads (Δ PL, Δ PR) to provide physiologic perfusion and balance left-right-sided flow rates. A model-based approach with extended Kalman and Golay-Savitzky filters was used to estimate Δ PL and Δ PR. Efficacy and robustness of the algorithm were evaluated in silico during simulated rest and exercise test conditions for: 1) excessive Δ PL and/or Δ PR setpoints; 2) rapid threefold increase in pulmonary vascular or vena caval resistances; 3) transitions from exercise to rest; and 4) ventricular fibrillation. Results and Conclusion: The proposed sensorless BiVAD algorithm successfully prevented suction, restored physiologic perfusion, and inherently maintained left-right-sided balance for all test conditions. Significance: The proposed algorithm does not require any device modification and may be integrated into current clinical BiVADs.Keywords
Funding Information
- National Heart, Lung, and Blood Institute (1R15HL115556-01A1)
- University of Louisville
This publication has 38 references indexed in Scilit:
- Comparison of Preload‐Sensitive Pressure and Flow Controller Strategies for a Dual Device Biventricular Support SystemArtificial Organs, 2011
- Biventricular Assist Devices: A Technical ReviewAnnals of Biomedical Engineering, 2011
- Response of Rotary Blood Pumps to Changes in Preload and Afterload at a Fixed Speed Setting Are Unphysiological When Compared With the Natural HeartArtificial Organs, 2011
- A Passively Controlled Biventricular Support DeviceArtificial Organs, 2010
- Optimizing the Response From a Passively Controlled Biventricular Assist DeviceArtificial Organs, 2010
- An innovative, sensorless, pulsatile, continuous-flow total artificial heart: Device design and initial in vitro studyThe Journal of Heart and Lung Transplantation, 2010
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceNew England Journal of Medicine, 2009
- Use of Cardiac Allografts With Mild and Moderate Left Ventricular Hypertrophy Can Be Safely Used in Heart Transplantation to Expand the Donor PoolJACC, 2008
- Dynamic Systemic Vascular Resistance in a Sheep Supported with a Nimbus AxiPumpASAIO Journal, 1994
- Determination of right ventricular wall thickness in systole and diastole. Echocardiographic and necropsy correlation in 32 patients.Heart, 1978