Outcomes of manual versus remote magnetic navigation for catheter ablation of ventricular tachycardia: a systematic review and updated meta‐analysis
- 6 April 2021
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 44 (6) , 1102-1114
Abstract
Purpose: Ventricular tachycardia (VT) ablation is a complex procedure that requires remarkable catheter manipulation skill, great mapping accuracy and catheter stability, and can expose patients to serious complications. Magnetic navigation system (RMN)‐guided ablation and contact force‐sensing (CFS) catheters have the potential to overcome these obstacles. We performed a systematic review and updated meta‐analysis of all available studies evaluating the outcomes of VT ablation by using RMN‐guided compared to manual navigation (MAN)‐guided, with and without CFS catheters.Methods: MEDLINE/PubMed, Cochrane, and Google Scholar were searched for randomized controlled trials (RCT) or observational studies with multivariate adjustment comparing RMN‐guided versus MAN‐guided VT ablation.Results: Thirteen studies enrolling 1348 patients (656 RMN‐guided vs. 692 MAN‐guided) were included. CFS catheter were used in 14% of MAN‐guided patients. In comparison to MAN‐guided and CFS‐guided, RMN‐guided VT ablation was associated with a significant higher acute ablation success (OR 2.32, 1.66–3.23 and OR 2.91, 1.29–6.53, respectively) but similar results in term of long‐term VT recurrence (OR 0.75, 0.56–1.01 and OR 0.79, 0.27–2.36, respectively). RMN‐guided showed a better safety profile (for all complications, OR 0.52, 0.34–0.81) and allowed a significant x‐ray reduction compared to MAN‐guided (OR 0.21, 0.14–0.32) and CFS‐guided VT ablation (OR 0.23, 0.11–0.52, all 95% CI).Conclusions: RMN‐guided was superior to MAN‐guided and CFS‐guided VT ablation in term of acute ablation success, all complications endpoint, and reduction of fluoroscopy exposure, but did not reduce long‐term VT recurrence. Large prospective multicenter randomized trials are needed to confirm these findings.Keywords
This publication has 25 references indexed in Scilit:
- Outcomes of repeat catheter ablation using magnetic navigation or conventional ablationEP Europace, 2013
- Catheter Ablation of Ventricular Tachycardias Using Remote Magnetic Navigation: A Consecutive Case–Control StudyJournal of Cardiovascular Electrophysiology, 2012
- Long‐Term Efficacy of Single Procedure Remote Magnetic Catheter Navigation for Ablation of Ischemic Ventricular Tachycardia: A Retrospective StudyJournal of Cardiovascular Electrophysiology, 2012
- Safety and efficacy of the remote magnetic navigation for ablation of ventricular tachycardias—a systematic reviewJournal of Interventional Cardiac Electrophysiology, 2011
- The magnetic navigation system allows safety and high efficacy for ablation of arrhythmiasEP Europace, 2011
- Endo-epicardial ablation of ventricular arrhythmias in the left ventricle with the Remote Magnetic Navigation System and the 3.5-mm open irrigated magnetic catheter: Results from a large single-center case–control seriesHeart Rhythm, 2010
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- Utilization of the PICO framework to improve searching PubMed for clinical questionsBMC Medical Informatics and Decision Making, 2007
- RF Catheter Ablation: Lessons on LesionsPacing and Clinical Electrophysiology, 2006
- Initial Experience With Remote Catheter Ablation Using a Novel Magnetic Navigation SystemCirculation, 2004