The effective point of measurement for depth‐dose measurements in small MV photon beams with different detectors
Authors:
Open Access
- 28 August 2019
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 46 (11) , 5209-5215
Abstract
Purpose: The effective point of measurement (EPOM) of cylindrical ionization chambers differs from their geometric center. The exact shift depends on chamber construction details, above all the chamber size, and to some degree on the field‐size and beam quality. It generally decreases as the chamber dimensions get smaller. In this work, effective points of measurement in small photon fields of a range of cylindrical chambers of different sizes are investigated, including small chambers that have not been studied previously.Methods: In this investigation, effective points of measurement for different ionization chambers (Farmer type, scanning chambers, micro‐ionization chambers) and solid state detectors were determined by measuring depth‐ionization curves in a 6 MV beam in field sizes between 2 × 2 cm2 and 10 × 10 cm2 and comparing those curves with curves measured with plane‐parallel chambers.Results: It was possible to average the results to one shift per detector, as the results were sufficiently independent of the studied field sizes. For cylindrical ion chambers, shifts of the EPOM were determined to be between 0.49 and 0.30 times the inner chamber radius from the reference point.Conclusions: We experimentally confirmed the previously reported decrease of the EPOM shift with decreasing detector size. Highly accurate data for a large range of detectors, including new very small ones, were determined. Thus, small chambers noticeably differ from the 0.5‐times to 0.6‐times the inner chamber radius recommendations in current dosimetry protocols. The detector‐individual EPOMs need to be considered for measurements of depth‐dose curves.Keywords
This publication has 20 references indexed in Scilit:
- Extraction of depth-dependent perturbation factors for silicon diodes using a plastic scintillation detectorMedical Physics, 2011
- Experimental determination of the effective point of measurement for various detectors used in photon and electron beam dosimetryPhysics in Medicine & Biology, 2011
- Multichannel film dosimetry with nonuniformity correctionMedical Physics, 2011
- Extraction of depth‐dependent perturbation factors for parallel‐plate chambers in electron beams using a plastic scintillation detectorMedical Physics, 2010
- Effective point of measurement of thimble ion chambers in megavoltage photon beamsMedical Physics, 2009
- Functional representation of tissue phantom ratios for photon fieldsMedical Physics, 2009
- Fast modelling of spectra and stopping-power ratios using differentiated fluence pencil kernelsPhysics in Medicine & Biology, 2008
- Measurement of output factors for small photon beamsMedical Physics, 2007
- Determination of the recombination correction factorkSfor some specific plane-parallel and cylindrical ionization chambers in pulsed photon and electron beamsPhysics in Medicine & Biology, 2006
- Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beamsMedical Physics, 2005