Reduction in sample injection bias using pressure gradients generated on chip
Authors:
- 10 February 2021
- journal article
- research article
- Published by Wiley in Electrophoresis
- Vol. 42 (7-8) , 983-990
Abstract
Sample injection in microchip‐based capillary zone electrophoresis (CZE) frequently rely on the use of electric fields which can introduce differences in the injected volume for the various analytes depending on their electrophoretic mobilities and molecular diffusivities. While such injection biases may be minimized by employing hydrodynamic flows during the injection process, this approach typically requires excellent dynamic control over the pressure gradients applied within a microfluidic network. The current article describes a microchip device that offers this needed control by generating pressure gradients on‐chip via electrokinetic means to minimize the dead volume in the system. In order to realize the desired pressure‐generation capability, an electric field was applied across two channel segments of different depths to produce a mismatch in the electroosmotic flow rate at their junction. The resulting pressure‐driven flow was then utilized to introduce sample zones into a CZE channel with minimal injection bias. The reported injection strategy allowed the introduction of narrow sample plugs with spatial standard deviations down to about 45 μm. This injection technique was later integrated to a capillary zone electrophoresis process for analyzing amino acid samples yielding separation resolutions of about 4–6 for the analyte peaks in a 3 cm long analysis channel.Keywords
Funding Information
- National Science Foundation (CHE‐1808507)
This publication has 57 references indexed in Scilit:
- Microfluidic flow counterbalanced capillary electrophoresisThe Analyst, 2013
- Sample introduction techniques for microchip electrophoresis: A reviewAnalytica Chimica Acta, 2012
- Hydrodynamic injection with pneumatic valving for microchip electrophoresis with total analyte utilizationElectrophoresis, 2011
- Integration of on‐chip peristaltic pumps and injection valves with microchip electrophoresis and electrochemical detectionElectrophoresis, 2010
- A low molecular weight cut-off polymer–silicate membrane for microfluidic applicationsMicrofluidics and Nanofluidics, 2010
- Generation of Nonbiased Hydrodynamic Injections on Microfluidic Devices Using Integrated Dielectric Elastomer ActuatorsAnalytical Chemistry, 2009
- Integration of serpentine channels for microchip electrophoresis with a palladium decoupler and electrochemical detectionElectrophoresis, 2009
- A micro-electrophoresis system based on a short capillary with hydrostatic pressure assisted separation and injectionMicrochimica Acta, 2009
- Electrokinetic‐based injection modes for separative microsystemsElectrophoresis, 2007
- Study of injection bias in a simple hydrodynamic injection in microchip CEElectrophoresis, 2007