Decreased imaging time of amyloid PET using [18F]florbetapir can maintain quantitative accuracy
Authors:
- 3 March 2022
- journal article
- research article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 15 (2) , 116-124
Abstract
Shortening the amount of time required to acquire amyloid positron emission tomography (PET) brain images while maintaining the accuracy of quantitative evaluation would help to overcome motion artifacts associated with Alzheimer’s disease patients. The present study aimed to validate the quantitative accuracy of [18F]florbetapir ([18F]FBP) imaging over a shorter acquisition duration. Forty participants were injected with [18F]FBP, and PET images were acquired for 50–55, 50–60, and 50–70 min after injection. Three physicians visually assessed the reprocessed [18F]FBP images using a binary scale to classify them as amyloid β (Aβ) negative or positive. A mean composite standard uptake value ratio (cSUVR) > 1.075 was defined as Aβ-positive based on receiver operating characteristic curves. Inter-reader and inter-acquisition duration agreements with visual assessment were evaluated using Cohen’s kappa (κ). Binary visual discrimination of 102 for the 120 [18F]FBP images, was consistent among the three readers. Sixteen, sixteen, and fourteen of the 40 [18F]FBP images acquired for 50–55, 50–60, and 50–70 min after injection, respectively, were deemed Aβ-positive by visual assessment. The inter-rater agreement was high, and the inter-acquisition duration agreement was almost perfect. The cSUVR did not change significantly among the acquisition durations, and the acquisition duration did not affect the outcome of discrimination based on the cSUVR cutoff. A shorter acquisition duration changed the visual assessment outcomes. Stable quantitative values were derived from [18F]FBP images acquired within 5 min. cSUVR helped to improve the performance and confidence in the outcomes of visual assessment.Keywords
Funding Information
- Japan Society for the Promotion of Science (JP20K16747)
- Eli Lilly Japan
This publication has 38 references indexed in Scilit:
- Regional Amyloid Deposition in Amnestic Mild Cognitive Impairment and Alzheimer's Disease Evaluated by [18F]AV-45 Positron Emission Tomography in Chinese PopulationPLOS ONE, 2013
- Amyloid-β Imaging with Pittsburgh Compound B and Florbetapir: Comparing Radiotracers and Quantification MethodsJournal of Nuclear Medicine, 2012
- Head motion evaluation and correction for PET scans with 18F-FDG in the Japanese Alzheimer’s disease neuroimaging initiative (J-ADNI) multi-center studyAnnals of Nuclear Medicine, 2012
- Performance Characteristics of Amyloid PET with Florbetapir F 18 in Patients with Alzheimer's Disease and Cognitively Normal SubjectsJournal of Nuclear Medicine, 2012
- Using PET with 18F-AV-45 (florbetapir) to quantify brain amyloid load in a clinical environmentEuropean Journal of Nuclear Medicine and Molecular Imaging, 2012
- Using Positron Emission Tomography and Florbetapir F 18 to Image Cortical Amyloid in Patients With Mild Cognitive Impairment or Dementia Due to Alzheimer DiseaseArchives of Neurology, 2011
- Positron Emission Tomography-Computed Tomography Standardized Uptake Values in Clinical Practice and Assessing Response to TherapySeminars in Ultrasound, CT and MRI, 2010
- In Vivo Imaging of Amyloid Deposition in Alzheimer Disease Using the Radioligand 18F-AV-45 (Flobetapir F 18)Journal of Nuclear Medicine, 2010
- Off-line motion correction methods for multi-frame PET dataEuropean Journal of Nuclear Medicine and Molecular Imaging, 2009
- Imaging brain amyloid in Alzheimer's disease with Pittsburgh Compound‐BAnnals of Neurology, 2004