Enhanced analysis of gating latency in 0.35T MR‐linac through innovative time synchronization of a motion phantom and plastic scintillation detector
Open Access
- 13 May 2025
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 26 (7) , e70116
Abstract
Purpose: This study aims to evaluate how different gantry angles, breathing rates (BPM), cine image speeds, and tracking algorithms affect beam on/off latency and the subsequent impact on target dose for a 0.35T MR‐Linac with a 6 MV FFF beam. The investigation incorporates an image‐based MRI4D modus QA motion phantom (MQA) and a measurement‐based plastic scintillation detector (PSD). Methods: The MQA's target was customized with an insertion for a 1 mm PSD from BluePhysics. Both the PSD and the MQA were simultaneously synchronized to the Linac to capture latency signals. A plan was created in the ViewRay TPS to deliver dose to the target at three gantry angles (0°, 120°, and 240°). Each gantry angle was evaluated at three breathing rates (10, 12, and 15 BPM). The study also examined two imaging speeds (4 and 8 FPS) and four tracking algorithms. Results: Across all configurations at 4 FPS, the overall mean beam‐on latency was 0.339 ± 0.06 s from the PSD and 0.318 ± 0.06 s from the MQA, whereas at 8 FPS it was 0.630 ± 0.07 s (PSD) and 0.609 ± 0.07 s (MQA). Conversely, the overall mean beam‐off latency at 4 FPS was 0.153 ± 0.03 s (PSD) and 0.124 ± 0.03 s (MQA), while at 8 FPS it was 0.121 ± 0.06 s (PSD) and 0.205 ± 0.04 s (MQA). The overall mean difference between gating and non‐gating doses was an increase of 12.050 ± 9.2 cGy at 4 FPS and 14.044 ± 7.4 cGy at 8 FPS. Conclusion: This comprehensive study underscores the significant influence of gantry angle, breathing rate, cine imaging speed, and tracking algorithms on latency and dose delivery accuracy in a 0.35T MR‐Linac.Keywords
This publication has 58 references indexed in Scilit:
- Film measurement and analytical approach for assessing treatment accuracy and latency in a magnetic resonance‐guided radiotherapy systemJournal of Applied Clinical Medical Physics, 2023
- Dosimetric evaluation of respiratory gating on a 0.35‐T magnetic resonance–guided radiotherapy linacJournal of Applied Clinical Medical Physics, 2022
- Clinical experience of MRI4D QUASAR motion phantom for latency measurements in 0.35T MR‐LINACJournal of Applied Clinical Medical Physics, 2020
- First experimental investigation of simultaneously tracking two independently moving targets on an MRI‐linac using real‐time MRI and MLC trackingMedical Physics, 2020
- Commissioning of a 1.5T Elekta Unity MR‐linac: A single institution experienceJournal of Applied Clinical Medical Physics, 2020
- MRIgRT dynamic lung motion thorax anthropomorphic QA phantom: Design, development, reproducibility, and feasibility studyMedical Physics, 2019
- Using adaptive magnetic resonance image‐guided radiation therapy for treatment of inoperable pancreatic cancerCancer Medicine, 2019
- AlignRT® and Catalyst™ in whole‐breast radiotherapy with DIBH: Is IGRT still needed?Journal of Applied Clinical Medical Physics, 2019
- First clinical implementation of real‐time, real anatomy tracking and radiation beam controlMedical Physics, 2018
- Dosimetric validation of a magnetic resonance image gated radiotherapy system using a motion phantom and radiochromic filmJournal of Applied Clinical Medical Physics, 2017