A preclinical radiotherapy dosimetry audit using a realistic 3D printed murine phantom
Open Access
- 26 April 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-13
Abstract
Preclinical radiation research lacks standardized dosimetry procedures that provide traceability to a primary standard. Consequently, ensuring accuracy and reproducibility between studies is challenging. Using 3D printed murine phantoms we undertook a dosimetry audit of Xstrahl Small Animal Radiation Research Platforms (SARRPs) installed at 7 UK centres. The geometrically realistic phantom accommodated alanine pellets and Gafchromic EBT3 film for simultaneous measurement of the dose delivered and the dose distribution within a 2D plane, respectively. Two irradiation scenarios were developed: (1) a 10 × 10 mm2 static field targeting the pelvis, and (2) a 5 × 5 mm2 90° arc targeting the brain. For static fields, the absolute difference between the planned dose and alanine measurement across all centres was 4.1 ± 4.3% (mean ± standard deviation), with an overall range of − 2.3 to 10.5%. For arc fields, the difference was − 1.2% ± 6.1%, with a range of − 13.1 to 7.7%. EBT3 dose measurements were greater than alanine by 2.0 ± 2.5% and 3.5 ± 6.0% (mean ± standard deviation) for the static and arc fields, respectively. 2D dose distributions showed discrepancies to the planned dose at the field edges. The audit demonstrates that further work on preclinical radiotherapy quality assurance processes is merited.Keywords
Funding Information
- National Centre for the Replacement Refinement and Reduction of Animals in Research (NC/P00203X/1)
- Cancer Research UK (C147/A25254, C1994/A28701)
- NIHR Manchester Biomedical Research Centre
This publication has 51 references indexed in Scilit:
- A national dosimetry audit of intraoperative radiotherapyBritish Journal of Radiology, 2013
- The Importance of Dosimetry Standardization in RadiobiologyJournal of Research of the National Institute of Standards and Technology, 2013
- Simulation of realistic linac motion improves the accuracy of a Monte Carlo based VMAT plan QA systemRadiotherapy and Oncology, 2013
- Marvin: an anatomical phantom for dosimetric evaluation of complex radiotherapy of the head and neckPhysics in Medicine & Biology, 2013
- Establishing a Murine Model of the Hematopoietic Syndrome of the Acute Radiation SyndromeHealth Physics, 2012
- MOSFET Assessment of Radiation Dose Delivered to Mice Using the Small Animal Radiation Research Platform (SARRP)Radiation Research, 2011
- Specific issues in small animal dosimetry and irradiator calibrationInternational Journal of Radiation Biology, 2011
- A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platformPhysics in Medicine & Biology, 2009
- Realistic reference phantoms: An ICRP/ICRU joint effortAnnals of the ICRP, 2009
- High-Resolution, Small Animal Radiation Research Platform With X-Ray Tomographic Guidance CapabilitiesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008