Analytical methods for system matrix calculation and spatial resolution evaluation of DC-SPECT system
- 18 July 2025
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (14) , 145031
Abstract
Objective. To develop and evaluate an analytical method for calculating the system matrix of a dynamic cardiac single photon emission computed tomography (DC-SPECT) system, eliminating the need for computationally intensive Monte Carlo (MC) simulations. Approach. An analytical model was proposed for system matrix generation, incorporating solid angle calculations adapted for square-shaped pinhole collimators. The resulting sensitivity maps were validated against MC simulation results. Image reconstructions using the proposed analytical model were compared to those using an MC-simulated system matrix. Additionally, the overall system performance, including sensitivity and spatial resolution, was assessed using MC simulations. Main results. The analytical sensitivity map showed good agreement with MC-based results. Reconstructed images using the analytical model preserved key features but showed reduced performance compared to MC-based reconstructions. MC simulations of DC-SPECT demonstrated a spatial resolution of 5.5 mm for hot regions and 6.0 mm for cold regions, with an overall system sensitivity of 0.07% over a 15 cm diameter spherical field of view. Significance. The proposed analytical method offers a fast and practical alternative to MC-based system matrix generation, enabling efficient system design and prototyping. While its use in clinical image reconstruction requires further evaluation, the model provides a promising tool for accelerating the development of next-generation DC-SPECT systems.Keywords
Funding Information
- Norges Idrettshøgskole (R01HL145160)
This publication has 45 references indexed in Scilit:
- Advances in Clinical Care with Contemporary Cardiac SPECTJournal of Medical Imaging and Radiation Sciences, 2024
- SPECT at the speed of PET: a feasibility study of CZT-based whole-body SPECT/CT in the post 177Lu-DOTATATE and 177Lu-PSMA617 settingEuropean Journal of Nuclear Medicine and Molecular Imaging, 2023
- Recent Developments in SPECT/CTSeminars in Nuclear Medicine, 2022
- Two decades of SPECT/CT – the coming of age of a technology: An updated review of literature evidenceEuropean Journal of Nuclear Medicine and Molecular Imaging, 2019
- Advances in imaging instrumentation for nuclear cardiologyJournal of Nuclear Cardiology, 2017
- Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM)European Journal of Nuclear Medicine and Molecular Imaging, 2016
- New Cardiac Cameras: Single-Photon Emission CT and PETSeminars in Nuclear Medicine, 2014
- 3D iteratively reconstructed spatial resolution map and sensitivity characterization of a dedicated cardiac SPECT cameraJournal of Nuclear Cardiology, 2014
- The origins of SPECT and SPECT/CTEuropean Journal of Nuclear Medicine and Molecular Imaging, 2013
- Advances in SPECT camera software and hardware: Currently available and new on the horizonJournal of Nuclear Cardiology, 2012