Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator
Authors:
Open Access
- 10 January 2021
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 22 (2) , 165-177
Abstract
Purpose: A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C‐SPECT, and a NaI scintillation detector‐equipped Anger‐type SPECT (A‐SPECT) scanner, with a LEHR and LMEGP.Methods: A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99mTc and 123I. We also measured the system's sensitivity and high‐contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high‐contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction.Results: The energy resolution with 99mTc was 5.6% in C‐SPECT with WEHR and 9.9% in A‐SPECT with LEHR. Using 123I, the results were 9.1% in C‐SPECT with WEHR, 5.5% in C‐SPECT with MEHRS, and 10.4% in A‐SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C‐SPECT performed better in SPECT acquisition. High‐contrast resolution was improved in C‐SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C‐SPECT.Conclusion: C‐SPECT demonstrates excellent energy resolution and improved high‐contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C‐SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high‐contrast resolution.Keywords
This publication has 28 references indexed in Scilit:
- Optimization of iterative reconstruction parameters with 3-dimensional resolution recovery, scatter and attenuation correction in 123I-FP-CIT SPECTAnnals of Nuclear Medicine, 2015
- Scatter correction improves concordance in SPECT MPI with a dedicated cardiac SPECT solid-state cameraJournal of Nuclear Cardiology, 2015
- Model-based correction for scatter and tailing effects in simultaneous99mTc and123I imaging for a CdZnTe cardiac SPECT cameraPhysics in Medicine & Biology, 2015
- Factors affecting the myocardial activity acquired during exercise SPECT with a high-sensitivity cardiac CZT camera as compared with conventional Anger cameraEuropean Journal of Nuclear Medicine and Molecular Imaging, 2013
- Quantification of 99mTc-DPD concentration in the lumbar spine with SPECT/CTEJNMMI Research, 2013
- Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom modelsRadiological Physics and Technology, 2012
- Cardiac Dedicated Ultrafast SPECT Cameras: New Designs and Clinical ImplicationsJournal of Nuclear Medicine, 2011
- Development of an ultra-high resolution SPECT system with a CdTe semiconductor detectorAnnals of Nuclear Medicine, 2009
- A Novel High-Sensitivity Rapid-Acquisition Single-Photon Cardiac Imaging CameraJournal of Nuclear Medicine, 2009
- Recent progress in CdTe and CdZnTe detectorsIEEE Transactions on Nuclear Science, 2001