Failure of a Femoral Revision Knee Component with Chemical Corrosion and Elevated Metal Ions
- 1 July 2023
- journal article
- review article
- Published by Wolters Kluwer Health in JBJS Case Connector
- Vol. 13 (3)
Abstract
Case: Failure of the modular junction in revision total knee arthroplasty is a rare complication. We report a patient with late, atraumatic failure of a modern, modular revision femoral component, with preoperative elevation of serum cobalt and chromium levels. Retrieval analysis showed extensive chemical corrosion. Conclusion: Failure of a modern, modular femoral component may cause metal synovitis and elevated serum metal levels. Subtle radiographic changes and preoperative serum metal levels may identify this complication.Keywords
This publication has 10 references indexed in Scilit:
- Fatigue Failure of Semiconstrained Total Knee Replacement: Magnetic Resonance Imaging Diagnosis With Tips and Tricks for Extraction and ReconstructionArthroplasty Today, 2022
- Nontribological corrosion modes dominate wrought CoCrMo acetabular taper corrosion: A retrieval studyJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2021
- Modular junction fractures in a modern rotating-platform knee arthroplasty systemArthroplasty Today, 2019
- Trunnion corrosion in metal-on-polyethylene hip arthroplastyThe Bone & Joint Journal, 2018
- Mechanical, chemical and biological damage modes within head‐neck tapers of CoCrMo and Ti6Al4V contemporary hip replacementsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2017
- Grain Boundary Assisted Crevice Corrosion in CoCrMo AlloysCorrosion, 2016
- Metallosis following Knee Arthroplasty: A Histological and Immunohistochemical StudyInternational Journal of Immunopathology and Pharmacology, 2011
- Failure at the Taper Lock of a Modular Stemmed Femoral Implant in Revision Knee ArthroplastyJournal of Bone and Joint Surgery, 2007
- Failure of the stem–condyle junction of a modular femoral stem in revision total knee arthroplastyThe Journal of Arthroplasty, 2001
- Disengagement of a Locking Screw from a Modular Stem in Revision Total Knee Arthroplasty. A Report of Three Cases*Journal of Bone and Joint Surgery, 1997