Comparison of Near-Infrared Oximeters in a Liquid Optical Phantom with Varying Intralipid and Blood Content
- 1 January 2016
- book chapter
- Published by Springer Nature
- Vol. 876, 413-418
Abstract
The interpretation of cerebral tissue oxygen saturation values (StO2) in clinical settings is currently complicated by the use of different near-infrared spectrophotometry (NIRS) devices producing different StO2 values for the same oxygenation due to differences in the algorithms and technical aspects. The aim was to investigate the effect of changes in scattering and absorption on the StO2 of different NIRS devices in a liquid optical phantom. We compared three continuous-wave (CW) with a frequency domain (FD) NIRS device. Responsiveness to oxygenation changes was only slightly altered by different intralipid (IL) concentrations. However, alterations in haematocrit (htc) showed a strong effect: increased htc led to a 20–35 % increased response of all CW devices compared to the FD device, probably due to differences in algorithms regarding the water concentration.Keywords
This publication has 13 references indexed in Scilit:
- Cerebral oxygenation after birth – a comparison of INVOS ® and FORE ‐ SIGHT ™ near‐infrared spectroscopy oximetersActa Paediatrica, 2014
- Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy VolunteersAnesthesia & Analgesia, 2013
- Peripheral tissue oximetry: comparing three commercial near-infrared spectroscopy oximeters on the forearmJournal of Clinical Monitoring and Computing, 2013
- Calibration of a prototype NIRS oximeter against two commercial devices on a blood-lipid phantomBiomedical Optics Express, 2013
- The Effect of Basic Assumptions on the Tissue Oxygen Saturation Value of Near Infrared Spectroscopy2012
- A Review of near Infrared Spectroscopy for Term and Preterm NewbornsJournal of Near Infrared Spectroscopy, 2012
- Has the time come to use near-infrared spectroscopy as a routine clinical tool in preterm infants undergoing intensive care?Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences, 2011
- Intralipid: towards a diffusive reference standard for optical tissue phantomsPhysics in Medicine & Biology, 2010
- Comparison of cerebral oxygen saturation in premature infants by near-infrared spatially resolved spectroscopy: observations on probe-dependent biasJournal of Biomedical Optics, 2008
- Performance Comparison of Several Published Tissue Near-Infrared Spectroscopy AlgorithmsAnalytical Biochemistry, 1995