The reduction of 176 Lu background in Lu-based PET scanners using optimized classification
- 28 August 2020
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 65 (17) , 175016
Abstract
Positron emission tomography (PET) using scanners incorporating lutetium-based (Lu-based) scintillators are widely used in nuclear medicine. However their application in imaging very low (99%). It was subsequently tested by imaging point-like source, planar Derenzo and bar phantoms with the pair of TOF-PET detectors. An 89.4% image contrast enhancement for the Derenzo phantom at an activity concentration of 100 Bq mm −2 , and a 52.4% peak-to-valley ratio improvement across the area of bar phantom at a concentration of 25 Bq mm −2 , were observed on the reconstructed images with XGBoost classification applied. The proposed method could extend the usage of Lu-based PET scanners to very low activity detection and imaging and has the potential to be used in a variety of molecular imaging tasks to detect low-level signals.Keywords
Funding Information
- National Institutes of Health (R35 CA197608)
This publication has 21 references indexed in Scilit:
- Evaluation of a BGO-Based PET System for Single-Cell Tracking Performance by Simulation and Phantom StudiesMolecular Imaging, 2016
- 90Y ‐PET imaging: Exploring limitations and accuracy under conditions of low counts and high random fractionMedical Physics, 2015
- Simultaneous Whole-Body PET/MR Imaging in Comparison to PET/CT in Pediatric Oncology: Initial ResultsRadiology, 2014
- Reduction method for intrinsic random coincidence events from 176Lu in low activity PET imagingRadiological Physics and Technology, 2014
- Validation of a GATE Model of $^{176}$Lu Intrinsic Radioactivity in LSO PET SystemsIEEE Transactions on Nuclear Science, 2011
- Lutetium oxyorthosilicate (LSO) intrinsic activity correction and minimal detectable target activity study for SPECT imaging with a LSO-based animal PET scannerPhysics in Medicine & Biology, 2008
- CT, MRI, PET, PET/CT, and Ultrasound in the Evaluation of Obstetric and Gynecologic PatientsSurgical Clinics of North America, 2008
- GATE: a simulation toolkit for PET and SPECTPhysics in Medicine & Biology, 2004
- Greedy function approximation: A gradient boosting machine.The Annals of Statistics, 2001
- In-beam PET imaging for the control of heavy-ion tumour therapyIEEE Transactions on Nuclear Science, 1997