Abstract
Positron emission tomography (PET) using scanners incorporating lutetium-based (Lu-based) scintillators are widely used in nuclear medicine. However their application in imaging very low (99%). It was subsequently tested by imaging point-like source, planar Derenzo and bar phantoms with the pair of TOF-PET detectors. An 89.4% image contrast enhancement for the Derenzo phantom at an activity concentration of 100 Bq mm −2 , and a 52.4% peak-to-valley ratio improvement across the area of bar phantom at a concentration of 25 Bq mm −2 , were observed on the reconstructed images with XGBoost classification applied. The proposed method could extend the usage of Lu-based PET scanners to very low activity detection and imaging and has the potential to be used in a variety of molecular imaging tasks to detect low-level signals.
Funding Information
  • National Institutes of Health (R35 CA197608)

This publication has 21 references indexed in Scilit: