Demonstration of record sensitivity for water vapor detection by means of comb-locked cavity ring-down spectroscopy
- 9 September 2024
- journal article
- research article
- Published by Optica Publishing Group in Optica
- Vol. 11 (9) , 1277-1284
Abstract
We report on the development, characterization, and test of a comb-locked cavity ring-down spectrometer (CL-CRDS) operating in the spectral region around 1.39 µm. The system is based on the use of a hemispherical optical resonator with a finesse as high as ∼507000, which gives an empty-cavity ring-down time of about 285 µs. An Allan-Werle analysis on repeated acquisitions of the ring-down time at a fixed laser frequency suggests a minimum detectable absorption coefficient of 2×10−12cm−1 for the optimum integration time of 45 s. This limit can be exceeded by adopting the strategy of long-term spectral averaging. Taking advantage of the frequency stability guaranteed by the optical frequency comb, the CL-CRDS spectra were averaged over more than two days, thus removing efficiently the effect of mechanical, acoustic, and thermal noises. As a result, we could achieve a minimum detectable absorption coefficient as low as 3.7×10−13cm−1, which corresponds to a limit of detection for H2O in N2 of nine parts per trillion and a H2O partial pressure of 2×10−8 Pa (or 2×10−10 mbar). The potentialities of our approach are demonstrated by recording the absorption features of HD16O and HD18O in flows of ultra-high-purity N2 and ambient air, respectively.Keywords
Funding Information
- European Metrology Programme for Innovation and Research (20IND06)
This publication has 39 references indexed in Scilit:
- Advances in cavity-enhanced methods for high precision molecular spectroscopy and test of fundamental physicsLa Rivista del Nuovo Cimento, 2024
- The HITRAN2020 molecular spectroscopic databaseJournal of Quantitative Spectroscopy and Radiative Transfer, 2022
- Calculated line lists for H216O and H218O with extensive comparisons to theoretical and experimental sources including the HITRAN2016 databaseJournal of Quantitative Spectroscopy and Radiative Transfer, 2020
- Recent advances in collisional effects on spectra of molecular gases and their practical consequencesJournal of Quantitative Spectroscopy and Radiative Transfer, 2018
- Room temperature line lists for deuterated waterJournal of Quantitative Spectroscopy and Radiative Transfer, 2017
- Development of a portable heavy-water leak sensor based on laser absorption spectroscopyAnnals of Nuclear Energy, 2016
- An isolated line-shape model to go beyond the Voigt profile in spectroscopic databases and radiative transfer codesJournal of Quantitative Spectroscopy and Radiative Transfer, 2013
- First detection of transitions in the second quadrupole overtone band of nitrogen near 1.44μm by CW-CRDS with 6×10−13cm−1 sensitivityChemical Physics Letters, 2013
- Advances in laser-based isotope ratio measurements: selected applicationsApplied Physics B, 2008
- Lineshapes of the 172 and 602 GHz rotational transitions of HC15NJournal of Molecular Spectroscopy, 2007