Characterization of high capacitance electrodes for the application of direct current electrical nerve block
- 11 September 2015
- journal article
- research article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 54 (1) , 191-203
Abstract
Direct current (DC) can briefly produce a reversible nerve conduction block in acute experiments. However, irreversible reactions at the electrode–tissue interface have prevented its use in both acute and chronic settings. A high capacitance material (platinum black) using a charge-balanced waveform was evaluated to determine whether brief DC block (13 s) could be achieved repeatedly (>100 cycles) without causing acute irreversible reduction in nerve conduction. Electrochemical techniques were used to characterize the electrodes to determine appropriate waveform parameters. In vivo experiments on DC motor conduction block of the rat sciatic nerve were performed to characterize the acute neural response to this novel nerve block system. Complete nerve motor conduction block of the rat sciatic nerve was possible in all experiments, with the block threshold ranging from −0.15 to −3.0 mA. DC pulses were applied for 100 cycles with no nerve conduction reduction in four of the six platinum black electrodes tested. However, two of the six electrodes exhibited irreversible conduction degradation despite charge delivery that was within the initial Q (capacitance) value of the electrode. Degradation of material properties occurred in all experiments, pointing to a possible cause of the reduction in nerve conduction in some platinum black experiments .Keywords
Funding Information
- National Institute of Neurological Disorders and Stroke (1R01NS074149)
This publication has 48 references indexed in Scilit:
- Investigating the electrochemical windows of ionic liquidsJournal of Industrial and Engineering Chemistry, 2013
- Separated interface nerve electrode prevents direct current induced nerve damageJournal of Neuroscience Methods, 2011
- Design, fabrication and evaluation of a conforming circumpolar peripheral nerve cuff electrode for acute experimental useJournal of Neuroscience Methods, 2011
- Conduction block of whole nerve without onset firing using combined high frequency and direct currentMedical & Biological Engineering & Computing, 2010
- Platinum Black Polymer Electrolyte Membrane Based Electrodes RevisitedJournal of the Electrochemical Society, 2008
- High‐frequency electrical conduction block of mammalian peripheral motor nerveMuscle & Nerve, 2005
- Direct current electrical conduction block of peripheral nerveIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2004
- Selective Detrusor Activation by Electrical Stimulation of the Human SacralNerve RootsArtificial Organs, 1997
- A model of safe levels for electrical stimulationIEEE Transactions on Biomedical Engineering, 1992
- Hydrogen Adsorption on Platinum Single Crystal ElectrodesJournal of the Electrochemical Society, 1965