Standardisation of conventional and advanced iterative reconstruction methods for Gallium-68 multi-centre PET-CT trials
Open Access
- 17 July 2021
- journal article
- research article
- Published by Springer Nature in EJNMMI Physics
- Vol. 8 (1) , 1-19
Abstract
Purpose: To assess the applicability of the Fluorine-18 performance specifications defined by EANM Research Ltd (EARL), in Gallium-68 multi-centre PET-CT trials using conventional (ordered subset expectation maximisation, OSEM) and advanced iterative reconstructions which include the systems’ point spread function (PSF) and a Bayesian penalised likelihood algorithm (BPL) commercially known as Q.CLEAR. The possibility of standardising the two advanced reconstruction methods was examined. Methods: The NEMA image quality phantom was filled with Gallium-68 and scanned on a GE PET-CT system. PSF and BPL with varying post-reconstruction Gaussian filter width (2–6.4 mm) and penalisation factor (200–1200), respectively, were applied. The average peak-to-valley ratio from six profiles across each sphere was estimated to inspect any edge artefacts. Image noise was assessed using background variability and image roughness. Six GE and Siemens PET-CT scanners provided Gallium-68 images of the NEMA phantom using both conventional and advanced reconstructions from which the maximum, mean and peak recoveries were drawn. Fourteen patients underwent 68Ga-PSMA PET-CT imaging. BPL (200-1200) reconstructions of the data were compared against PSF smoothed with a 6.4-mm Gaussian filter. Results: A Gaussian filter width of approximately 6 mm for PSF and a penalisation factor of 800 for BPL were needed to suppress the edge artefacts. In addition, those reconstructions provided the closest agreement between the two advanced iterative reconstructions and low noise levels with the background variability and the image roughness being lower than 7.5% and 11.5%, respectively. The recoveries for all methods generally performed at the lower limits of the EARL specifications, especially for the 13- and 10-mm spheres for which up to 27% (conventional) and 41% (advanced reconstructions) lower limits are suggested. The lesion standardised uptake values from the clinical data were significantly different between BPL and PSF smoothed with a Gaussian filter of 6.4 mm wide for all penalisation factors except for 800 and 1000. Conclusion: It is possible to standardise the advanced reconstruction methods with the reconstruction parameters being also sufficient for minimising the edge artefacts and noise in the images. For both conventional and advanced reconstructions, Gallium-68 specific recovery coefficient limits were required, especially for the smallest phantom spheres.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (WT 203148/Z/16/Z)
- Cancer Research UK
- Medical Research Council
- National Institute for Health Research
- NIHR Biomedical Research Centre at Guy's and St Thomas' NHS Foundation Trust and King's College London
This publication has 36 references indexed in Scilit:
- Resolution modeling in PET imaging: Theory, practice, benefits, and pitfallsMedical Physics, 2013
- Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patientsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2013
- PET imaging with a [68Ga]gallium-labelled PSMA ligand for the diagnosis of prostate cancer: biodistribution in humans and first evaluation of tumour lesionsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2012
- Improvement in PET/CT Image Quality with a Combination of Point-Spread Function and Time-of-Flight in Relation to Reconstruction ParametersJournal of Nuclear Medicine, 2012
- Positron range in PET imaging: an alternative approach for assessing and correcting the blurringPhysics in Medicine & Biology, 2012
- Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluationPhysics in Medicine & Biology, 2010
- From RECIST to PERCIST: Evolving Considerations for PET Response Criteria in Solid TumorsJournal of Nuclear Medicine, 2009
- The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trialsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2008
- Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parametersEuropean Journal of Nuclear Medicine and Molecular Imaging, 2006
- Analysis of Emission Tomographic Scan Data: Limitations Imposed by Resolution and BackgroundJournal of Computer Assisted Tomography, 1984