Proton beam dosimetry in the presence of magnetic fields using Farmer‐type ionization chambers of different radii
Open Access
- 29 March 2023
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 50 (7) , 4590-4599
Abstract
Background: Magnetic resonance‐guided proton therapy is promising, as it combines high‐contrast imaging of soft tissue with highly conformal dose delivery. However, proton dosimetry in magnetic fields using ionization chambers is challenging since the dose distribution as well as the detector response are perturbed.Purpose: This work investigates the effect of the magnetic field on the ionization chamber response, and on the polarity and ion recombination correction factors, which are essential for the implementation of a proton beam dosimetry protocol in the presence of magnetic fields.Methods: Three Farmer‐type cylindrical ionization chambers, the 30013 with 3 mm inner radius (PTW, Freiburg, Germany) and two custom built chambers “R1” and “R6” with 1 and 6 mm inner radii respectively were placed at the center of an experimental electromagnet (Schwarzbeck Mess – Elektronik, Germany) 2 cm depth of an in‐house developed 3D printed water phantom. The detector response was measured for a 3 × 10 cm2 field of mono‐energetic protons 221.05 MeV/u for the three chambers, and with an additional proton beam of 157.43 MeV/u for the chamber PTW 30013. The magnetic flux density was varied between 0.1 and 1.0 Tesla in steps of 0.1 Tesla.Results: At both energies, the ionization chamber PTW 30013 showed a non‐linear response as a function of the magnetic field strength, with a decrease of the ionization chamber response of up to 0.27% ± 0.06% (1 SD) at 0.2 Tesla, followed by a smaller effect at higher magnetic field strength. For the chamber R1, the response decreased slightly with the magnetic field strength up to 0.45% ± 0.12% at 1 Tesla, and for the chamber R6, the response decreased up to 0.54% ± 0.13% at 0.1 Tesla, followed by a plateau up to 0.3 Tesla, and a weaker effect at higher magnetic field strength. The dependence of the polarity and recombination correction factor on the magnetic field was ⩽0.1% for the chamber PTW 30013.Conclusions: The magnetic field has a small but significant effect on the chamber response in the low magnetic field region for the chamber PTW 30013 and for R6, and in the high magnetic field region for the chamber R1. Corrections may be necessary for ionization chamber measurements, depending on both the chamber volume and the magnetic flux density. No significant effect of the magnetic field on the polarity and recombination correction factor was detected in this work for the ionization chamber PTW 30013.Keywords
This publication has 10 references indexed in Scilit:
- A review on reference dosimetry in radiation therapy with proton and light ion beams: status and impact of new developmentsRadiation Measurements, 2022
- Integrated MRI‐guided proton therapy planning: Accounting for the full MRI field in a perpendicular systemMedical Physics, 2021
- Development and results of a patient-reported treatment experience questionnaire on a 1.5 T MR-LinacClinical and Translational Radiation Oncology, 2021
- MR‐guided proton therapy: Impact of magnetic fields on the detector responseMedical Physics, 2020
- Partial breast irradiation with the 1.5 T MR-Linac: First patient treatment and analysis of electron return and stream effectsRadiotherapy and Oncology, 2020
- First application of a high‐resolution silicon detector for proton beam Bragg peak detection in a 0.95 T magnetic fieldMedical Physics, 2019
- Characterization of EBT3 radiochromic films for dosimetry of proton beams in the presence of magnetic fieldsMedical Physics, 2019
- Technical Note: Experimental verification of magnetic field‐induced beam deflection and Bragg peak displacement for MR‐integrated proton therapyMedical Physics, 2018
- Dosimetric feasibility of real‐time MRI‐guided proton therapyMedical Physics, 2014
- Particle therapy at the Heidelberg Ion Therapy Center (HIT) – Integrated research-driven university-hospital-based radiation oncology service in Heidelberg, GermanyRadiotherapy and Oncology, 2010