The Biomechanical Properties of Pedicle Screw Fixation Combined With Trajectory Bone Cement Augmentation in Osteoporotic Vertebrae
- 1 March 2016
- journal article
- Published by Wolters Kluwer Health in Clinical Spine Surgery
- Vol. 29 (2) , 78-85
Abstract
The biomechanics of pedicle screw fixation combined with trajectory cement augmentation with various filling volumes were measured by pull-out, periodic antibending, and compression fatigue tests. To investigate the biomechanical properties of the pedicle screw fixation combined with trajectory bone cement (polymethylmethacrylate) augmentation in osteoporotic vertebrae and to explore the optimum filling volume of the bone cement. Pedicle screw fixation is considered to be the most effective posterior fixation method. The decrease of the bone mineral density apparently increases the fixation failure risk caused by screw loosening and displacement. Trajectory bone cement augmentation has been confirmed to be an effective method to increase the bone intensity and could markedly increase the stability of the fixation interface. Sixteen elderly cadaveric 1-5 lumbar vertebral specimens were diagnosed with osteoporosis. The left and right vertebral pedicles were alternatively randomized for treatment in all groups, with the contralateral pedicles as control. The study groups included: group A (pedicle screw fixation with full trajectory bone cement augmentation), group B (75% filling), group C (50% filling), and group D (25% filling). Finally, the bone cement leakage and dispersion were assessed and the mechanical testing was conducted. The bone cement was well dispersed around the pedicle screw. The augmented bone intensity, pull-out strength, periodic loading times, and compression fatigue performance were markedly higher than those of the control groups. With the increase in trajectory bone cement, the leakage was also increased (P<0.05). The pull-out strength of the pedicle screw was increased with an increase in bone mineral density and trajectory bone cement. It peaked at 75% filling, with the largest power consumption. The optimal filling volume of the bone cement was 75% of the trajectory volume (about 1.03 mL). The use of excessive bone cement did not increase the fixation intensity but increased the risk of leakage.Keywords
This publication has 16 references indexed in Scilit:
- Pedikelschraubenaugmentation aus biomechanischer SichtDer Orthopäde, 2010
- Polymethylmethacrylate-Augmented Screw Fixation for Stabilization of the Osteoporotic Spine : A Three-Year Follow-Up of 37 PatientsJournal of Korean Neurosurgical Society, 2009
- Assessment of different screw augmentation techniques and screw designs in osteoporotic spinesEuropean Spine Journal, 2008
- SEGMENTAL POLYMETHYLMETHACRYLATE-AUGMENTED PEDICLE SCREW FIXATION IN PATIENTS WITH BONE SOFTENING CAUSED BY OSTEOPOROSIS AND METASTATIC TUMOR INVOLVEMENTNeurosurgery, 2007
- Augmentation of a Loosened Sacral Pedicle Screw With Percutaneous Polymethylmethacrylate InjectionJournal of Spinal Disorders & Techniques, 2006
- Stability of Pedicle Screws After Kyphoplasty AugmentationJournal of Spinal Disorders & Techniques, 2006
- Pedicle screw fixation strength: pullout versus insertional torqueThe Spine Journal, 2004
- Internal Fracture Fixation in Patients With OsteoporosisJournal of the American Academy of Orthopaedic Surgeons, 2003
- Biomechanical Evaluation of a Double-Threaded Pedicle Screw in Elderly VertebraeJournal of Spinal Disorders & Techniques, 2002
- Complications Associated with Pedicle Screws*Journal of Bone and Joint Surgery, 1999