A new implementation of EN1839 method T mixture preparation to prepare calibration gas mixtures for anesthetic gas monitors
- 8 July 2020
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 35 (5) , 1069-1076
Abstract
In this study different mixtures from anesthetic liquids vapours in air and O2—sevoflurane/air, sevoflurane/O2 and isoflurane/O2—were prepared according to EN1839 method T. The gases mixtures were used to calibrate anesthetic gas monitors. The targeted concentrations (TCs) from sevoflurane and isoflurane in air and O2 were 0 vol%, 0.15 vol%, 0.7 vol%, 1 vol%, 1.5 vol%, 2.5 vol%, 5 vol%, 6 vol% and 8 vol%. The values of liquids, air and O2 flow rates which are corresponding theoretically to the TCs were calculated. The results showed that the measured concentrations by anesthetic gas monitor are in good agreement with TCs. Full validation was performed to assure the suitability of the method for preparing gas mixtures to calibrate anesthetic gas monitors. Linearity, limit of detection (LOD), limit of quantification (LOQ), precision, method robustness and uncertainty of measurement results were evaluated in the validation study. The results indicated that the method is valid for the purpose of analysis.Keywords
This publication has 14 references indexed in Scilit:
- Explosion regions of propane, isopropanol, acetone, and methyl acetate/inert gas/air mixturesJournal of Loss Prevention in the Process Industries, 2016
- Test Gas Generation from Pure Liquids: An Application-Oriented Overview of Methods in a NutshellInternational Journal of Chemical Engineering, 2012
- System for the generation of standard gas mixtures of volatile and semi-volatile organic compounds for calibrations of solid-phase microextraction and other sampling devicesJournal of Chromatography A, 2004
- Refractive indices for volatile anesthetic gases: Equipment and method for calibrating vaporizers and monitorsJournal of Clinical Monitoring and Computing, 1995
- Evaluation of three transportable multigas anesthetic monitors: The Bruel & Kjaer Anesthetic Gas Monitor 1304, the Datex Capnomac Ultima, and the Nellcor N-2500Journal of Clinical Monitoring and Computing, 1993
- Methods to produce calibration mixtures for anesthetic gas monitors and how to perform volumetric calculations on anesthetic gasesJournal of Clinical Monitoring and Computing, 1992
- Generation of standard gaseous mixtures by thermal decomposition of surface compoundsJournal of Chromatography A, 1991
- A gas mixer for computer calibration of an anesthetic mass spectrometerJournal of Clinical Monitoring and Computing, 1986
- Generation of standard gaseous mixturesJournal of Chromatography A, 1984
- The preparation of standard gas mixtures. A reviewThe Analyst, 1981