Surface and Buildup Region Dose Measurements with Markus Parallel-Plate Ionization Chamber, GafChromic EBT3 Film, and MOSFET Detector for High-Energy Photon Beams
Authors:
Open Access
- 12 June 2016
- journal article
- research article
- Published by Hindawi Limited in Advances in High Energy Physics
- Vol. 2016, 1-10
Abstract
The aim of the study was to investigate surface and buildup region doses for 6 MV and 15 MV photon beams using a Markus parallel-plate ionization chamber, GafChromic EBT3 film, and MOSFET detector for different field sizes and beam angles. The measurements were made in a water equivalent solid phantom at the surface and in the buildup region of the 6 MV and 15 MV photon beams at 100 cm source-detector distance for 5 × 5, 10 × 10, and 20 × 20 cm2field sizes and 0°, 30°, 60°, and 80° beam angles. The surface doses using 6 MV photon beams for 10 × 10 cm2field size were found to be 20.3%, 18.8%, and 25.5% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface doses using 15 MV photon beams for 10 × 10 cm2field size were found to be 14.9%, 13.4%, and 16.4% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface dose increased with field size for all dosimeters. As the angle of the incident radiation beam became more oblique, the surface dose increased. The effective measurement depths of dosimeters vary; thus, the results of the measurements could be different. This issue can lead to mistakes at surface and buildup dosimetry and must be taken into account.Keywords
All Related Versions
This publication has 24 references indexed in Scilit:
- Water equivalent properties of materials commonly used in proton dosimetryApplied Radiation and Isotopes, 2014
- Surface dose and build-up region measurements with wedge filters for 6 and 18 MV photon beamsJapanese Journal of Radiology, 2010
- Surface dose measurements with GafChromic EBT film for 6 and 18MV photon beamsPhysica Medica, 2009
- Accurate skin dose measurements using radiochromic film in clinical applicationsMedical Physics, 2006
- Radiochromic film for medical radiation dosimetryMaterials Science and Engineering: R: Reports, 2003
- Measurement of radiotherapy x‐ray skin dose on a chest wall phantomMedical Physics, 2000
- A new radiotherapy surface dose detector: The MOSFETMedical Physics, 1996
- Entrance and exit dose regions for a clinac-2100cInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1993
- Determination of build‐up region over‐response corrections for a Markus‐type chamberMedical Physics, 1990
- Radiochromic plastic films for accurate measurement of radiation absorbed dose and dose distributionsRadiation Physics and Chemistry (1977), 1977