The effect of MLC speed and acceleration on the plan delivery accuracy of VMAT
- 1 May 2015
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 88 (1049) , 20140698
Abstract
To determine a new metric utilizing multileaf collimator (MLC) speeds and accelerations to predict plan delivery accuracy of volumetric modulated arc therapy (VMAT). To verify VMAT delivery accuracy, gamma evaluations, analysis of mechanical parameter difference between plans and log files, and analysis of changes in dose-volumetric parameters between plans and plans reconstructed with log files were performed with 40 VMAT plans. The average proportion of leaf speeds ranging from l to h cm s(-1) (Sl-h and l-h = 0-0.4, 0.4-0.8, 0.8-1.2, 1.2-1.6 and 1.6-2.0), mean and standard deviation of MLC speeds were calculated for each VMAT plan. The same was carried out for accelerations in centimetre per second squared (Al-h and l-h = 0-4, 4-8, 8-12, 12-16 and 16-20). The correlations of those indicators to plan delivery accuracy were analysed with Spearman's correlation coefficient (rs). The S1.2-1.6 and mean acceleration of MLCs showed generally higher correlations to plan delivery accuracy than did others. The highest rs values were observed between S1.2-1.6 and global 1%/2 mm (rs = -0.698 with p < 0.001) as well as mean acceleration and global 1%/2 mm (rs = -0.650 with p < 0.001). As the proportion of MLC speeds and accelerations >0.4 and 4 cm s(-2) increased, the plan delivery accuracy of VMAT decreased. The variations in MLC speeds and accelerations showed considerable correlations to VMAT delivery accuracy. As the MLC speeds and accelerations increased, VMAT delivery accuracy reduced.Keywords
This publication has 33 references indexed in Scilit:
- An adaptive planning strategy for station parameter optimized radiation therapy (SPORT): Segmentally boosted VMATMedical Physics, 2013
- On the sensitivity of common gamma‐index evaluation methods to MLC misalignments in Rapidarc quality assuranceMedical Physics, 2013
- Dosimetric evaluation of Tomotherapy and four-box field conformal radiotherapy in locally advanced rectal cancerRadiation Oncology Journal, 2013
- Comparison of IMRT and VMAT Techniques in Spine Stereotactic Radiosurgery with International Spine Radiosurgery Consortium Consensus GuidelinesProgress in Medical Physics, 2013
- Implementation of phantom-less IMRT delivery verification using Varian DynaLog files and R/V outputPhysics in Medicine & Biology, 2012
- Quality assurance of dynamic parameters in volumetric modulated arc therapyBritish Journal of Radiology, 2012
- Assessing software upgrades, plan properties and patient geometry using intensity modulated radiation therapy (IMRT) complexity metricsMedical Physics, 2011
- Per‐beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errorsMedical Physics, 2011
- On the impact of dose rate variation upon RapidArc® implementation of volumetric modulated arc therapyMedical Physics, 2010
- What is an acceptably smoothed fluence? Dosimetric and delivery considerations for dynamic sliding window IMRTRadiation Oncology, 2007