Three-dimensional computational evaluation of water-to-air mass stopping power ratio for therapeutic helium-, carbon-, oxygen-, and neon-ion beams and its application to patient-specific quality assurance
- 13 November 2026
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 71 (21) , 215037
Abstract
Objective. Current dosimetry protocols recommend a constant water-to-air mass stopping power ratio, s w,air , of 1.126 with an uncertainty of ±1.5% for light ions, irrespective of ion species and energy. This study aims to develop a computational framework to calculate three-dimensional s w,air distributions of helium-, carbon-, oxygen-, and neon-ion beams for patient specific quality assurance (PSQA), and to systematically investigate spatial variations in s w,air under clinically relevant conditions. Approach. A computational framework was developed to calculate three-dimensional s w,air distributions based on the Bragg-Gray approximation, together with three-dimensional dose distributions for treatment planning. The framework was validated against published datasets for a mono-energetic carbon-ion beam. Treatment plans were generated using helium-, carbon-, oxygen-, and neon-ion beams for rectangular targets with varying spread-out Bragg peak (SOBP) widths and depths, as well as for bone and soft tissue, head and neck, and ocular cases. The corresponding s w,air distributions were calculated in a water phantom and compared with the recommended value of 1.126. Main Results. The mean s w,air within the target increased for shallower targets with narrower SOBP widths. For all ion species, s w,air increased toward the distal end of the SOBP, with larger variations observed for heavier ion species. In neon-ion irradiations for shallow targets, s w,air occasionally exceeded the ±1.5% uncertainty range in a limited region near the distal end of the SOBP. In all other cases, s w,air remained within the uncertainty range. Similar trends were observed in clinical cases. Based on these results, when considering the mean s w,air within the target, the suitable fixed values were 1.128 for carbon and oxygen ions and 1.129 for helium and neon ions. Significance . The proposed computational framework enables three-dimensional, plan-specific evaluation of s w,air and provides a practical approach for improving the dosimetric accuracy in charged-particle therapy with helium, carbon, oxygen, and neon ions, especially for PSQA.Keywords
Funding Information
- Japan Society for the Promotion of Science (24K10773)
- Exploratory Research for Advanced Technology (JPMJER2102)
This publication has 59 references indexed in Scilit:
- Monte Carlo–calculated beam–quality and perturbation correction factors for helium–, carbon–, and neon–ion beamsMedical Physics, 2025
- Possibility of increasing linear energy transfer in multi‐ion radiotherapy for bone and soft tissue sarcomasMedical Physics, 2025
- On the perturbation effect and LET dependence of beam quality correction factors in carbon ion beamsMedical Physics, 2022
- Monte Carlo computation of 3D distributions of stopping power ratios in light ion beam therapy using GATE‐RTionMedical Physics, 2021
- Development and Validation of Single Field Multi-Ion Particle Therapy TreatmentsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2020
- Dose detectors, sensors, and their applicationsMedical Physics, 2018
- Recent developments in Geant4Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Quality assurance of carbon ion and proton beams: A feasibility study for using the 2D MatriXX detectorPhysica Medica, 2016
- Magnetic scanning system for heavy ion therapyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993
- Depth-dose and stopping-power data for mono-energetic electron beamsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1990