In Vivo Evaluation of a Physiologic Control System for Rotary Blood Pumps Based on the Left Ventricular Pressure-Volume Loop
- 1 June 2022
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 68 (6) , 791-799
Abstract
Current generation continuous flow assist devices to operate at a fixed speed, which limits preload response and exercise capacity in left ventricular assist device (LVAD) patients. A feedback control system was developed to automatically adjust pump speed based on direct measurements of ventricular loading using a custom cannula tip with an integrated pressure sensor and volume-sensing conductance electrodes. The input to the control system is the integral of the left ventricular (LV) pressure versus conductance loop (PGA) over each cardiac cycle. The feedback control system adjusts pump speed based on the difference between the measured PGA and the desired PGA. The control system and cannula tip were tested in acute ovine studies (n = 5) using the HeartMate II LVAD. The preload response of the control system was evaluated by partially occluding and releasing the inferior vena cava using a vessel loop snare. The cannula tip was integrated onto a custom centrifugal flow LVAD and tested in a 14-day bovine study. The control system adjusted pump support to maintain constant ventricular loading: pump speed increased (decreased) following an increase (decrease) in preload. This study demonstrated in vivo the Starling-like response of an automatic pump control system based on direct measurements of LV loading.Keywords
This publication has 24 references indexed in Scilit:
- Changes in the Functional Status Measures of Heart Failure Patients With Mechanical Assist DevicesASAIO Journal, 2013
- Theoretical Foundations of a Starling‐Like Controller for Rotary Blood PumpsArtificial Organs, 2012
- Exercise Capacity in Patients Supported With Rotary Blood Pumps Is Improved by a Spontaneous Increase of Pump Flow at Constant Pump Speed and by a Rise in Native Cardiac OutputArtificial Organs, 2011
- Response of Rotary Blood Pumps to Changes in Preload and Afterload at a Fixed Speed Setting Are Unphysiological When Compared With the Natural HeartArtificial Organs, 2011
- Health-related quality of life and exercise tolerance in recipients of heart transplants and left ventricular assist devices: A prospective, comparative studyThe Journal of Heart and Lung Transplantation, 2011
- Comparison of Cardiac Power Output and Exercise Performance in Patients With Left Ventricular Assist Devices, Explanted (Recovered) Patients, and Those With Moderate to Severe Heart FailureThe American Journal of Cardiology, 2010
- Assessment of Arterial Blood Pressure During Support With an Axial Flow Left Ventricular Assist DeviceThe Journal of Heart and Lung Transplantation, 2009
- First Clinical Experience with an Automatic Control System for Rotary Blood Pumps During Ergometry and Right-Heart CatheterizationThe Journal of Heart and Lung Transplantation, 2006
- Single-Beat Estimation of End-Systolic Pressure-Volume Relation in HumansCirculation, 1996
- Linearity of the Frank-Starling relationship in the intact heart: the concept of preload recruitable stroke work.Circulation, 1985