Impact of machine log-files uncertainties on the quality assurance of proton pencil beam scanning treatment delivery
Authors:
- Simona ToscanoSTSimona Toscano
- Carles GomàCGCarles Gomà
- A Barragán-MonteroABA Barragán-Montero
- 29 April 2019
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 64 (9) , 095021
Abstract
Irradiation log-files store useful information about the plan delivery, and together with independent Monte Carlo dose engine calculations can be used to reduce the time needed for patient-specific quality assurance (PSQA). Nonetheless, machine log-files carry an uncertainty associated to the measurement of the spot position and intensity that can influence the correct evaluation of the quality of the treatment delivery. This work addresses the problem of the inclusion of these uncertainties for the final verification of the treatment delivery. Dedicated measurements performed in an IBA Proteus Plus gantry with a pencil beam scanning (PBS) dedicated nozzle have been carried out to build a ‘room-dependent’ model of the spot position uncertainties. The model has been obtained through interpolation of the look-up tables describing the systematic and random uncertainties, and it has been tested for a clinical case of a brain cancer patient irradiated in a dry-run. The delivered dose has been compared with the planned dose with the inclusion of the errors obtained applying the model. Our results suggest that the accuracy of the treatment delivery is higher than the spot position uncertainties obtained from the log-file records. The comparison in terms of DVHs shows that the log-reconstructed dose is compatible with the planned dose within the 95% confidence interval obtained applying our model. The initial mean dose difference between the calculated dose to the patient based on the plan and recorded data is around 1%. The difference is essentially due to the log-file uncertainties and it can be removed with a correct treatment of these errors. In conclusion our new PSQA protocol allows for a fast verification of the dose delivered after every treatment fraction through the use of machine log-files and an independent Monte Carlo dose engine. Moreover, the inclusion of log-file uncertainties in the dose calculation allows for a correct evaluation of the quality of the treatment plan delivery.Keywords
Funding Information
- Fonds Baillet-Latour
This publication has 18 references indexed in Scilit:
- Performance of a hybrid Monte Carlo‐Pencil Beam dose algorithm for proton therapy inverse planningMedical Physics, 2017
- Treatment log files as a tool to identify treatment plan sensitivity to inaccuracies in scanned proton beam deliveryRadiotherapy and Oncology, 2017
- Combined influence of CT random noise and HU-RSP calibration curve nonlinearities on proton range systematic errorsPhysics in Medicine & Biology, 2017
- Reference dosimetry of proton pencil beams based on dose-area product: a proof of conceptPhysics in Medicine & Biology, 2017
- Evaluation of motion mitigation using abdominal compression in the clinical implementation of pencil beam scanning proton therapy of liver tumorsMedical Physics, 2016
- Assessing the quality of proton PBS treatment delivery using machine log files: comprehensive analysis of clinical treatments delivered at PSI Gantry 2Physics in Medicine & Biology, 2016
- Independent dose calculations for commissioning, quality assurance and dose reconstruction of PBS proton therapyPhysics in Medicine & Biology, 2015
- Range uncertainties in proton therapy and the role of Monte Carlo simulationsPhysics in Medicine & Biology, 2012
- Intensity modulation methods for proton radiotherapyPhysics in Medicine & Biology, 1999
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998