Suppression Of Afterglow In CsI(Tl) By Codoping With Eu/sup 2+/: Fabrication Of Microcolumnar Films For High-Resolution High-Speed Imaging
Authors:
- V. V. NagarkarVNV. V. Nagarkar
- S. C. ThackerSTS. C. Thacker
- C. BrecherCBC. Brecher
- 1 January 2006
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 6 (10823654) , 3567-3570
Abstract
Although CsI:Tl is one of the brightest and most efficient scintillator materials ever developed, its applicability for high speed imaging is limited by its persistent afterglow. We have demonstrated that much of this afterglow can be suppressed by the addition of Eu2+ to the CsI host lattice. The magnitude of the effect depends strongly on the duration of the excitation pulse, but is virtually independent of its intensity. Our observations indicate that the presence of Eu2+ ions introduces a set of electron traps that fundamentally alter the decay kinetics, making the material useful for such applications as high speed radiography and, perhaps, X-ray CT. At RMD several single crystals of CsI:Tl,Eu with Eu2+ concentrations between 0.05 to 0.5 mole percent have been grown using the Bridgman technique. Here we report on our fabrication of microcolumnar films of the material using thermal vapor deposition for high-resolution digital radiography applications. Our data suggests that the afterglow suppression effect has a profound impact on high speed imaging performance, allowing rapid data acquisition with substantially reduced persistence from previous frames. The latest results on the effect of the codopant on the luminescence from CsI:Tl crystals as well as microcolumnar films will be reported.Keywords
This publication has 7 references indexed in Scilit:
- Suppression of afterglow in CsI:Tl by codoping with Eu2+—I: ExperimentalNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- Suppression of afterglow in CsI:Tl by codoping with Eu2+—II: Theoretical modelNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- Mammographic imaging with a small format CCD‐based digital cassette: Physical characteristics of a clinical systemMedical Physics, 2000
- Full breast digital mammography with an amorphous silicon‐based flat panel detector: Physical characteristics of a clinical prototypeMedical Physics, 2000
- A ghost story: Spatio-temporal response characteristics of an indirect-detection flat-panel imagerMedical Physics, 1999
- DQE(f) of four generations of computed radiography acquisition devicesMedical Physics, 1995
- A simple method for determining the modulation transfer function in digital radiographyIEEE Transactions on Medical Imaging, 1992