Self-Monitoring of Blood Glucose: The Use of the First or the Second Drop of Blood
Open Access
- 17 February 2011
- journal article
- Published by American Diabetes Association in Diabetes Care
- Vol. 34 (3) , 556-560
Abstract
OBJECTIVE: There is no general agreement regarding the use of the first or second drop of blood for glucose monitoring. This study investigated whether capillary glucose concentrations, as measured in the first and second drops of blood, differed ≥10% compared with a control glucose concentration in different situations.RESEARCH DESIGN AND METHODS: Capillary glucose concentrations were measured in two consecutive drops of blood in the following circumstances in 123 patients with diabetes: without washing hands, after exposing the hands to fruit, after washing the fruit-exposed hands, and during application of different amounts of external pressure around the finger. The results were compared with control measurements.RESULTS: Not washing hands led to a difference in glucose concentration of ≥10% in the first and in the second drops of blood in 11% and 4% of the participants, respectively. In fruit-exposed fingers, these differences were found in 88% and 11% of the participants, respectively. Different external pressures led to ≥10% differences in glucose concentrations in 5–13% of the participants.CONCLUSIONS: We recommend washing the hands with soap and water, drying them, and using the first drop of blood for self-monitoring of blood glucose. If washing hands is not possible, and they are not visibly soiled or exposed to a sugar-containing product, it is acceptable to use the second drop of blood after wiping away the first drop. External pressure may lead to unreliable readings.Keywords
This publication has 11 references indexed in Scilit:
- The influence of a soiled finger in capillary blood glucose monitoring.2010
- SKML-Quality Mark for point-of-care test (POCT) glucose meters and glucose meters for home-useClinical Chemistry and Laboratory Medicine (CCLM), 2010
- Blood glucose monitoring: Milking the finger and using the first drop of blood give correct glucose valuesDiabetes Research and Clinical Practice, 2009
- Improving the Quality of Self-Monitoring Blood Glucose Measurement: A Study in Reducing Calibration ErrorsDiabetes Technology & Therapeutics, 2006
- Quality Specifications for Glucose Meters: Assessment by Simulation Modeling of Errors in Insulin DoseClinical Chemistry, 2001
- Patient-derived Quality Specifications for Instruments Used in Self-Monitoring of Blood GlucoseClinical Chemistry, 2001
- Identifying Variables Associated With Inaccurate Self-Monitoring of Blood Glucose: Proposed Guidelines to Improve AccuracyThe Diabetes Educator, 2000
- Desirable standards for laboratory tests if they are to fulfill medical needsClinical Chemistry, 1993
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986