On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans
Open Access
- 6 March 2014
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 15 (2) , 67-84
Abstract
Partial transmission through rounded leaf ends of Varian multileaf collimators (MLC) is accounted for with a parameter called the dosimetric leaf gap (DLG). Verification of the value of the DLG is needed when the dose delivery is accompanied by gantry rotation in VMAT plans. We compared the doses measured with GAFCHROMIC film and an ionization chamber to treatment planning system (TPS) calculations to identify the optimum values of the DLG in clinical plans of the whole brain with metastases transferred to a phantom. We noticed the absence of a single value of the DLG that properly models all VMAT plans in our cohort (the optimum DLG varied between and ). The former value is considerably different from the optimum DLG in sliding window plans (about 2.0 mm) that approximate IMRT plans. We further found that a single-value DLG model cannot accurately reproduce the measured dose profile even of a uniform static slit at a fixed gantry, which is the simplest MLC-delimited field. The calculation overestimates the measurement in the proximal penumbra, while it underestimates in the distal penumbra. This prompted us to expand the DLG parameter from a plan-specific number to a mathematical concept of the DLG being a function of the distance in the beam's eye view (BEV) between the dose point and the leaf ends. Such function compensates for the difference between the penumbras in a beam delimited with a rounded leaf MLC and delimited with solid jaws. Utilization of this concept allowed us generating a pair of step-and-shoot MLC plans for which we could qualitatively predict the value of the DLG providing best match to ionization chamber measurements. The plan for which the leafs stayed predominantly at positions requiring low values of the DLG (as seen in the profiles of 1D slits) yielded the combined DLG of , while the plan with leafs staying at positions requiring larger values of the DLG yielded the DLG . Considering the DLG to be a function of the distance (in BEV) between the dose point and the leaf ends allowed us to provide an explanation as to why conventional single-number DLG is plan-specific in VMAT plans. PACS numbers: 87.56.jf, 87.56.nk
Keywords
This publication has 26 references indexed in Scilit:
- Verification and dosimetric impact of Acuros XB algorithm on intensity modulated stereotactic radiotherapy for locally persistent nasopharyngeal carcinomaMedical Physics, 2012
- Implementing RapidArc into clinical routine: A comprehensive program from machine QA to TPS validation and patient QAMedical Physics, 2011
- Evaluation of a new VMAT QA device, or the “X” and “O” array geometriesJournal of Applied Clinical Medical Physics, 2011
- Quality assurance of volumetric modulated arc therapy: Evaluation and comparison of different dosimetric systemsMedical Physics, 2011
- A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBTMedical Physics, 2010
- Effective point of measurement of thimble ion chambers in megavoltage photon beamsMedical Physics, 2009
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Commissioning and Quality Assurance of RapidArc Radiotherapy Delivery SystemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Volumetric modulated arc therapy: IMRT in a single gantry arcMedical Physics, 2007
- A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations*Physics in Medicine & Biology, 2005