Radiation dose optimisation in dynamic volume CT of the heart: tube current adaptation based on anterior–posterior chest diameter
- 27 April 2010
- journal article
- research article
- Published by Springer Nature in The International Journal of Cardiovascular Imaging
- Vol. 26 (8) , 933-940
Abstract
To compare tube current adaptation based on 3 body mass index (BMI) categories versus anterior-posterior chest diameter (APD) for radiation dose optimisation in patients undergoing dynamic volume cardiac CT. Two cardiac imaging centres participated in the study. 20 patients underwent a prospectively triggered 320-slice single beat cardiac CT using the X-ray tube current [mA] manually adjusted to the patient’s BMI (group I). In 20 subsequent patients, the tube current was adapted according to the patient’s APD (group II). All other parameters were kept constant. Image noise was defined as the standard deviation of attenuation values and measured using a ROI in the descending aorta. Variation in image noise was statistically compared between both patient groups. Average and standard deviation of pixel noise were 29.1 HU and 14.8 HU in group I and 28.0 HU and 4.2 HU in group II. Inter-individual variation of pixel noise was significantly lower in group II compared to group I (p < 0.0001). Tube current adaptation based on APD is superior to stepwise adaptation based on BMI for optimising radiation dose in dynamic volume cardiac CT and therefore limits unnecessary radiation dose while ensuring diagnostic image quality in patients with diverse body habitus.Keywords
This publication has 24 references indexed in Scilit:
- Diagnostic Accuracy of CT Coronary AngiographyCardiology Clinics, 2009
- Evaluation of Attenuation-Based Tube Current Control in Coronary Artery Calcium Scoring on Prospective ECG-triggered 64-detector CTAcademic Radiology, 2009
- Significance of Cardiac Computed Tomography Incidental Findings in Acute Chest PainThe American Journal of Medicine, 2009
- Radiation dose exposure of computed tomography coronary angiography: comparison of dual-source, 16-slice and 64-slice CTHeart, 2009
- Radiation Dose in Cardiac CTAmerican Journal of Roentgenology, 2009
- Patient‐specific dose estimation for pediatric chest CTMedical Physics, 2008
- Optimizing radiation dose and image qualityEuropean Radiology Supplements, 2007
- Simulator-assisted Setting of Scan Protocols for X-ray CT: Development and Clinical Usefulness of the Scan Plan SimulatorJapanese Journal of Radiological Technology, 2006
- Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studiesMedical Physics, 1999
- Body mass index as a measure of body fatness: age- and sex-specific prediction formulasBritish Journal of Nutrition, 1991