CT of the medial clavicular epiphysis for forensic age estimation: hands up?
Open Access
- 24 February 2021
- journal article
- research article
- Published by Springer Nature in International journal of legal medicine
- Vol. 135 (4) , 1581-1587
Abstract
Purpose: The aim of this study was to assess the impact of arm position in computed tomography (CT) of the clavicle performed for forensic age estimation on clavicular position, image noise, and radiation dose. Methods and materials: Forty-seven CT scans of the medial clavicular epiphysis performed for forensic age estimation were conducted with either hands and arms held upwards (CTHU, 28 persons) or positioned at the body (CTHD, 19 persons). Presets were identical for both positions (70 mAs/140 kVp; Brilliance iCT, Philips). Each CT scan was reconstructed with an iterative algorithm (i-Dose 4) and evaluated at the middle of the sternoclavicular joint. Clavicular angle was measured on a.p. topograms in relation to a horizontal line. Quantitative image noise was measured in air at the level of medial clavicular epiphysis. Effective dose and scan length were recorded. Results: Hands-up position compared with hands-down position resulted in a lower lateral body diameter (CTHU 41.1 ± 3.6 cm vs. CTHD 44.6 ± 3.1 cm; P = 0.03), a reduced quantitative image noise (CTHU: 39.5 ± 9.2; CTHD: 46.2 ± 8.3; P = 0.02), and lower CTDIvol (5.1 ± 1.4 mGy vs. 6.7 ± 1.8 mGy; P = 0.001). Scan length was longer in patients examined with hands up (HU: 8.5 ± 3.4 cm; HD: 6.2 ± 2.1 cm; P = 0.006). Mean effective dose for CTHU was 0.79 ± 0.32 mSv compared with 0.95 ± 0.38 mSv in CTHD (P = 0.12). Clavicular angle was 17° ± 6° in patients with hands down and 32° ± 7° in patients with hands up (P < 0.001). Conclusion: By elevated arm positioning, the image quality of clavicular CT scans can be improved while maintaining radiation dose compared with hands down. Clavicular position differs according to the hand position. Thus, positioning patients with elevated hands is advisable for forensic clavicular CT examinations, but multiplanar CT reconstructions should be adjusted to clavicular position and scan length should be reduced to a minimum.Keywords
This publication has 23 references indexed in Scilit:
- Dose-optimized computed tomography of the cervical spine in patients with shoulder pull-down: Is image quality comparable with a standard dose protocol in an emergency setting?European Journal of Radiology, 2019
- Forensic age estimation using computed tomography of the medial clavicular epiphysis: a systematic reviewInternational journal of legal medicine, 2018
- The effects of patient positioning when interpreting CT dose metrics: A phantom studyMedical Physics, 2017
- Pediatric Exposures to Ionizing Radiation: Carcinogenic ConsiderationsInternational Journal of Environmental Research and Public Health, 2016
- Forensic Age Estimation: Methods, Certainty, and the LawDeutsches Ärzteblatt international, 2016
- Möglichkeiten der Strahlenreduktion bei der CT des KörperstammsRadiologie up2date, 2014
- The value of sub-stages and thin slices for the assessment of the medial clavicular epiphysis: a prospective multi-center CT studyForensic Science, Medicine and Pathology, 2013
- Arm Raising at Exposure-controlled Multidetector Trauma CT of Thoracoabdominal Region: Higher Image Quality, Lower Radiation DoseRadiology, 2008
- Criteria for age estimation in living individualsInternational journal of legal medicine, 2008
- The Current State of Forensic Age Estimation of Live Subjects for the Purpose of Criminal ProsecutionForensic Science, Medicine and Pathology, 2005