Quantifying inadvertent thermal bowel injury from the monopolar instrument
- 29 April 2016
- journal article
- research article
- Published by Springer Nature in Surgical Endoscopy
- Vol. 30 (11) , 4776-4784
Abstract
Background Insulation defects are observed in 3–39 % of laparoscopic instruments. Electrosurgical injuries due to insulation defects or capacitive coupling remain an issue in laparoscopic surgery with a prevalence of 0.6–5 per thousand cases. Shielded instruments with active electrode monitoring (AEM) have been postulated to prevent these injuries. The benefit of these instruments has not been quantified. Most bowel injuries are unrecognized intra-operatively. Injury is revealed only after the patient exhibits peritonitis symptoms and surgical intervention to repair the bowel is required. These injuries may result in devastating and costly complications or mortality. The extent of bowel injury possible with commonly used generator settings and associated energy output has never been histologically defined. Our objectives in this experimental study were: quantify and compare the energy released through insulation defects or capacitive coupling with standard unshielded monopolar versus shielded instruments with (AEM), determine energy required to cause a visible burn, and relate the histological burn depth to a given amount of energy. Methods Ex vivo porcine jejunum was used for tissue testing. An oscilloscope measured energy output from three common electrosurgical generators at recommended power settings with standard or AEM instruments with insulation defects and in capacitive coupling scenarios. Presence of a visible burn was noted, and depth of tissue damage for a given amount of energy was measured histologically. Results All samples that received ≥3.8 J of energy had visible burns. As little as 10 J caused full wall thickness burns. 3.8 J was exceeded at the 30- and 50-W power settings in every experimental scenario using standard monopolar instruments; AEM instruments never approached this much energy. Conclusions Serious burn injury results from small amounts of energy leaked from standard instruments. AEM instruments appeared protective and did not leak sufficient energy to cause burn injuries to the bowel.Keywords
Funding Information
- Unrestricted grant from Encision
- National Institutes of Health (UL1 TR001082.)
This publication has 15 references indexed in Scilit:
- Ten years of progress—improved hysterectomy outcomes in Finland 1996–2006: a longitudinal observation studyBMJ Open, 2013
- Accuracy of Small-Intestine Contrast Ultrasonography, Compared With Computed Tomography Enteroclysis, in Characterizing Lesions in Patients With Crohn's DiseaseClinical Gastroenterology and Hepatology, 2013
- Monopolar Electrosurgery through Single-Port Laparoscopy: A Potential Hidden Hazard for Bowel BurnsJournal of Minimally Invasive Gynecology, 2011
- Insulation failure in robotic and laparoscopic instrumentation: a prospective evaluationAmerican Journal of Obstetrics and Gynecology, 2011
- Surgeon-controlled Factors That Reduce Monopolar Electrosurgery Capacitive Coupling During LaparoscopySurgical Laparoscopy, Endoscopy & Percutaneous Techniques, 2010
- Insulation failure in laparoscopic instrumentsSurgical Endoscopy, 2009
- Reproducibility of bowel ultrasonography in the evaluation of Crohn's diseaseDigestive and Liver Disease, 2008
- Laparoscopic instrument insulation failure: The hidden hazardJournal of Minimally Invasive Gynecology, 2007
- Laparoscopic-assisted vaginal hysterectomy: American Association of Gynecologic Laparoscopists' 1995 Membership SurveyThe Journal of the American Association of Gynecologic Laparoscopists, 1997
- Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgeryThe American Journal of Surgery, 1992