Adaptation of continuous‐flow cavity ring‐down spectroscopy for batch analysis of δ 13 C of CO 2 and comparison with isotope ratio mass spectrometry
- 18 July 2011
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 25 (16) , 2355-2360
Abstract
Measurements of δ 13 C in CO 2 have traditionally relied on samples stored in sealed vessels and subsequently analyzed using magnetic sector isotope ratio mass spectrometry (IRMS), an accurate but expensive and high‐maintenance analytical method. Recent developments in optical spectroscopy have yielded instruments that can measure δ 13 CO 2 in continuous streams of air with precision and accuracy approaching those of IRMS, but at a fraction of the cost. However, continuous sampling is unsuited for certain applications, creating a need for conversion of these instruments for batch operation. Here, we present a flask (syringe) adaptor that allows the collection and storage of small aliquots (20–30 mL air) for injection into the cavity ring‐down spectroscopy (CRDS) instrument. We demonstrate that the adaptor's precision is similar to that of traditional IRMS (standard deviation of 0.3‰ for 385 ppm CO 2 standard gas). In addition, the concentration precision (±0.3% of sample concentration) was higher for CRDS than for IRMS (±7% of sample concentration). Using the adaptor in conjunction with CRDS, we sampled soil chambers and found that soil‐respired δ 13 C varied between two different locations in a piñon‐juniper woodland. In a second experiment, we found no significant discrimination between the respiration of a small beetle (~5 mm) and its diet. Our work shows that the CRDS system is flexible enough to be used for the analysis of batch samples as well as for continuous sampling. This flexibility broadens the range of applications for which CRDS has the potential to replace magnetic sector IRMS. Copyright © 2011 John Wiley & Sons, Ltd.Keywords
Funding Information
- Los Alamos National Lab's Institute for Geophysics and Planetary Physics Minigrant Program (11-10329)
This publication has 21 references indexed in Scilit:
- A dynamic soil chamber system coupled with a tunable diode laser for online measurements of δ 13 C, δ 18 O, and efflux rate of soil‐respired CO 2Rapid Communications in Mass Spectrometry, 2010
- Analysis of low‐concentration gas samples with continuous‐flow isotope ratio mass spectrometry: eliminating sources of contamination to achieve high precisionRapid Communications in Mass Spectrometry, 2009
- Proof‐of‐concept study on the suitability of 13C‐urea as a marker substance for assessment of in vivo behaviour of oral colon‐targeted dosage formsBritish Journal of Pharmacology, 2009
- Testing Lagrangian atmospheric dispersion modelling to monitor CO2 and CH4 leakage from geosequestrationAtmospheric Environment, 2009
- A laboratory comparison of two methods used to estimate the isotopic composition of soil δ 13 CO 2 efflux at steady stateRapid Communications in Mass Spectrometry, 2008
- Uncertainties in static closed chamber measurements of the carbon isotopic ratio of soil-respired COSoil Biology and Biochemistry, 2005
- Seasonal cycle of carbon dioxide and its isotopic composition in an urban atmosphere: Anthropogenic and biogenic effectsJournal of Geophysical Research: Atmospheres, 2003
- Temporal variation in ?13C of ecosystem respiration in the Pacific Northwest: links to moisture stressOecologia, 2003
- 13C content of ecosystem respiration is linked to precipitation and vapor pressure deficitOecologia, 2002
- Using septum‐capped vials with continuous‐flow isotope ratio mass spectrometric analysis of atmospheric CO 2 for Keeling plot applicationsRapid Communications in Mass Spectrometry, 2001