Mapping gradient nonlinearity and miscalibration using diffusion‐weighted MR images of a uniform isotropic phantom
Open Access
- 4 August 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 86 (6) , 3259-3273
Abstract
Purpose: To use diffusion measurements to map the spatial dependence of the magnetic field produced by the gradient coils of an MRI scanner with sufficient accuracy to correct errors in quantitative diffusion MRI (DMRI) caused by gradient nonlinearity and gradient amplifier miscalibration.Theory and Methods: The field produced by the gradient coils is expanded in regular solid harmonics. The expansion coefficients are found by fitting a model to a minimum set of diffusion‐weighted images of an isotropic diffusion phantom. The accuracy of the resulting gradient coil field maps is evaluated by using them to compute corrected b‐matrices that are then used to process a multi‐shell diffusion tensor imaging (DTI) dataset with 32 diffusion directions per shell.Results: The method substantially reduces both the spatial inhomogeneity of the computed mean diffusivities (MD) and the computed values of the fractional anisotropy (FA), as well as virtually eliminating any artifactual directional bias in the tensor field secondary to gradient nonlinearity. When a small scaling miscalibration was purposely introduced in the x, y, and z, the method accurately detected the amount of miscalibration on each gradient axis.Conclusion: The method presented detects and corrects the effects of gradient nonlinearity and gradient gain miscalibration using a simple isotropic diffusion phantom. The correction would improve the accuracy of DMRI measurements in the brain and other organs for both DTI and higher order diffusion analysis. In particular, it would allow calibration of MRI systems, improving data harmony in multicenter studies.Keywords
Funding Information
- National Institute of Biomedical Imaging and Bioengineering (Intramural)
This publication has 31 references indexed in Scilit:
- Improved correction for gradient nonlinearity effects in diffusion‐weighted imagingJournal of Magnetic Resonance Imaging, 2012
- The use of MRI apparent diffusion coefficient (ADC) in monitoring the development of brain infarctionBMC Medical Imaging, 2011
- From single‐pulsed field gradient to double‐pulsed field gradient MR: gleaning new microstructural information and developing new forms of contrast in MRINMR in Biomedicine, 2010
- Diffusion Tensor Imaging of the BrainNeurotherapeutics, 2007
- Diffusional kurtosis imaging: The quantification of non‐gaussian water diffusion by means of magnetic resonance imagingMagnetic Resonance in Medicine, 2005
- Analysis and generalized correction of the effect of spatial gradient field distortions in diffusion‐weighted imagingMagnetic Resonance in Medicine, 2003
- Optimal strategies for measuring diffusion in anisotropic systems by magnetic resonance imagingMagnetic Resonance in Medicine, 1999
- Optimised diffusion-weighting for measurement of apparent diffusion coefficient (ADC) in human brainMagnetic Resonance Imaging, 1997
- A simple method for obtaining cross‐term‐free images for diffusion anisotropy studies in NMR microimagingMagnetic Resonance in Medicine, 1991
- NMR microscopy of dynamic displacements: k-space and q-space imagingJournal of Physics E: Scientific Instruments, 1988