Reducing cardiac implantable electronic device–induced artefacts in cardiac magnetic resonance imaging
Open Access
- 27 August 2022
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 33 (2) , 1229-1242
Abstract
Objectives: Cardiac implantable electronic device (CIED)–induced metal artefacts possibly significantly diminish the diagnostic value of magnetic resonance imaging (MRI), particularly cardiac MR (CMR). Right-sided generator implantation, wideband late-gadolinium enhancement (LGE) technique and raising the ipsilateral arm to the generator during CMR scanning may reduce the CIED-induced image artefacts. We assessed the impact of generator location and the arm-raised imaging position on the CIED-induced artefacts in CMR. Methods: We included all clinically indicated CMRs performed on patients with normal cardiac anatomy and a permanent CIED with endocardial pacing leads between November 2011 and October 2019 in our institution (n = 171). We analysed cine and LGE sequences using the American Heart Association 17-segment model for the presence of artefacts. Results: Right-sided generator implantation and arm-raised imaging associated with a significantly increased number of artefact-free segments. In patients with a right-sided pacemaker, the median percentage of artefact-free segments in short-axis balanced steady-state free precession LGE was 93.8% (IQR 9.4%, n = 53) compared with 78.1% (IQR 20.3%, n = 58) for left-sided pacemaker (p < 0.001). In patients with a left-sided implantable cardioverter-defibrillator, the median percentage of artefact-free segments reached 87.5% (IQR 6.3%, n = 9) using arm-raised imaging, which fell to 62.5% (IQR 34.4%, n = 9) using arm-down imaging in spoiled gradient echo short-axis cine (p = 0.02). Conclusions: Arm-raised imaging represents a straightforward method to reduce CMR artefacts in patients with left-sided generators and can be used alongside other image quality improvement methods. Right-sided generator implantation could be considered in CIED patients requiring subsequent CMR imaging to ensure sufficient image quality. Key Points: • Cardiac implantable electronic device (CIED)–induced metal artefacts may significantly diminish the diagnostic value of an MRI, particularly in cardiac MRIs. • Raising the ipsilateral arm relative to the CIED generator is a cost-free, straightforward method to significantly reduce CIED-induced artefacts on cardiac MRIs in patients with a left-sided generator. • Right-sided generator implantation reduces artefacts compared with left-sided implantation and could be considered in CIED patients requiring subsequent cardiac MRIs to ensure adequate image quality in the future.Keywords
Funding Information
- Helsingin ja Uudenmaan Sairaanhoitopiiri
- Ida Montinin Säätiö
- Suomen Radiologiyhdistys
This publication has 48 references indexed in Scilit:
- ACR guidance document on MR safe practices: 2013Journal of Magnetic Resonance Imaging, 2013
- Quantitative Assessment of Artifacts on Cardiac Magnetic Resonance Imaging of Patients With Pacemakers and Implantable Cardioverter-DefibrillatorsCirculation: Cardiovascular Imaging, 2011
- Safety of serial MRI in patients with implantable cardioverter defibrillatorsHeart, 2011
- Improved 3‐Tesla cardiac cine imaging using widebandMagnetic Resonance in Medicine, 2010
- Wideband SSFP: Alternating repetition time balanced steady state free precession with increased band spacingMagnetic Resonance in Medicine, 2007
- Comparison of Defibrillation Efficacy and Survival Associated With Right Versus Left Pectoral Placement for Implantable DefibrillatorsThe American Journal of Cardiology, 2007
- Strategy for Safe Performance of Extrathoracic Magnetic Resonance Imaging at 1.5 Tesla in the Presence of Cardiac Pacemakers in Non–Pacemaker-Dependent PatientsCirculation, 2006
- Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging SafeCirculation, 2004
- Magnetic resonance imaging and cardiac pacemaker safety at 1.5-TeslaJACC, 2004
- Standardized Myocardial Segmentation and Nomenclature for Tomographic Imaging of the HeartCirculation, 2002