Investigation of Photoplethysmography Behind the Ear for Pulse Oximetry in Hypoxic Conditions with a Novel Device (SPYDR)
Open Access
- 4 April 2020
- journal article
- research article
- Published by MDPI AG in Biosensors
- Vol. 10 (4) , 34
Abstract
Photoplethysmography (PPG) is a valuable technique for noninvasively evaluating physiological parameters. However, traditional PPG devices have significant limitations in high-motion and low-perfusion environments. To overcome these limitations, we investigated the accuracy of a clinically novel PPG site using SPYDR®, a new PPG sensor suite, against arterial blood gas (ABG) measurements as well as other commercial PPG sensors at the finger and forehead in hypoxic environments. SPYDR utilizes a reflectance PPG sensor applied behind the ear, between the pinna and the hairline, on the mastoid process, above the sternocleidomastoid muscle, near the posterior auricular artery in a self-contained ear cup system. ABG revealed accuracy of SPYDR with a root mean square error of 2.61% at a 70–100% range, meeting FDA requirements for PPG sensor accuracy. Subjects were also instrumented with SPYDR, as well as finger and forehead PPG sensors, and pulse rate (PR) and oxygen saturation (SpO2) were measured and compared at various reduced oxygen profiles with a reduced oxygen breathing device (ROBD). SPYDR was shown to be as accurate as other sensors in reduced oxygen environments with a Pearson’s correlation >93% for PR and SpO2. In addition, SPYDR responded to changes in SpO2 up to 50 s faster than PPG measurements at the finger and forehead.Keywords
Funding Information
- United States Government (W15QKN-19-9-1012)
This publication has 44 references indexed in Scilit:
- Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitationsRespiratory Medicine, 2013
- Measuring Arterial Oxygenation in a High Altitude Field Environment: Comparing Portable Pulse Oximetry With Blood Gas AnalysisWilderness & Environmental Medicine, 2013
- Reducing motion artifacts in photoplethysmograms by using relative sensor motion: phantom studyJournal of Biomedical Optics, 2012
- Noninvasive Monitoring by PhotoplethysmographyClinics in Perinatology, 2012
- Combined photoplethysmographic monitoring of respiration rate and pulse: a comparison between different measurement sites in spontaneously breathing subjectsActa Anaesthesiologica Scandinavica, 2007
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- Photoplethysmography and its application in clinical physiological measurementPhysiological Measurement, 2007
- Forehead oximetry in critically ill patients: the case for a new monitoring siteRespiratory Care Clinics of North America, 2004
- Skin photoplethysmography — a reviewComputer Methods and Programs in Biomedicine, 1989
- Response of skin photoplethysmography, laser Doppler flowmetry and transcutaneous oxygen tensiometry to stenosis-induced reductions in limb blood flowEuropean Journal of Vascular Surgery, 1989