B1+$ {\mathrm{B}}_1^{+} $ mapping near metallic implants using turbo spin echo pulse sequences
- 13 March 2025
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 94 (2) , 554-570
Abstract
Purpose: To propose a B 1 + mapping technique for imaging of body parts containing metal hardware, based on magnitude images acquired with turbo spin echo (TSE) pulse sequences. Theory and Methods: To encode the underlying B 1 + , multiple (two to four) TSE image sets with various excitation and refocusing flip angles were acquired. To this end, the acquired signal intensities were matched to a database of simulated signals which was generated by solving the Bloch equations taking into account the exact sequence parameters. The retrieved B 1 + values were validated against gradient‐recalled and spin echo dual angle methods, as well as a vendor‐provided TurboFLASH‐based mapping sequence, in gel phantoms and human subjects without and with metal implants. Results: In the absence of metal, phantom experiments demonstrated excellent agreement between the proposed technique using three or four flip angle sets and reference dual angle methods. In human subjects without metal implants, the proposed technique with three or four flip angle sets showed excellent correlation with the spin echo dual angle method. In the presence of metal, both phantoms and human subjects revealed a narrow range of B 1 + estimation with the reference techniques, whereas the proposed technique successfully resolved B 1 + near the metal. In select cases, the technique was implemented in conjunction with multispectral metal artifact reduction sequences and successfully applied for B 1 + shimming. Conclusion: The proposed technique enables resolution of B 1 + values in regions near metal hardware, overcoming susceptibility‐related and narrow‐range limitations of standard mapping techniques.Keywords
This publication has 55 references indexed in Scilit:
- Radiofrequency transmit calibration: A multi‐center evaluation of vendor‐provided radiofrequency transmit mapping methodsMedical Physics, 2019
- B1 mapping for bias‐correction in quantitative T1 imaging of the brain at 3T using standard pulse sequencesJournal of Magnetic Resonance Imaging, 2017
- Extended phase graphs: Dephasing, RF pulses, and echoes ‐ pure and simpleJournal of Magnetic Resonance Imaging, 2014
- High‐resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DESPOT2Magnetic Resonance in Medicine, 2004
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- Correction forB1 andB0 variations in quantitativeT2 measurements using MRIMagnetic Resonance in Medicine, 2000
- Radiofrequency power calibration for magnetic resonance imaging using signal phase as indicatorMagnetic Resonance in Medicine, 1999
- Imaging of the active B1 field in vivoMagnetic Resonance in Medicine, 1996
- The rician distribution of noisy mri dataMagnetic Resonance in Medicine, 1995
- Nonsusceptibility artifacts due to metallic objects in MR imagingJournal of Magnetic Resonance Imaging, 1995