Objectives: To assess the validity of light-induced and laser-induced fluorescence devices compared to the visual-tactile method for detecting secondary caries around resin composite restorations. Materials and Methods: The study included 20 participants with 30 resin-composite restored teeth. Restorations’ margins were examined using three diagnostic methods: the visual-tactile method (FDI criteria), the light-induced fluorescence camera (VistaCam iX), and the laser-induced fluorescence device (DIAGNOdent pen), and the reference was visual inspection after removal of defective restorations. The validity of each method was evaluated. Inter-examiner reliability was calculated using Cohen’s kappa statistics. The level of significance was set at P = 0.05. Results: DIAGNOdent pen showed the highest sensitivity (100%) followed by VistaCam (98.82%) and the visual-tactile method (98.82%) at the enamel threshold. DIAGNOdent pen and VistaCam had lower specificity values than the visual-tactile method (81.69%, 76.06%, and 88.73% respectively). At the dentin threshold, DIAGNOdent pen yielded the highest sensitivity (89.36%), whereas VistaCam had the lowest (8.51%). The sensitivity of the visual-tactile method was low (57.45%) whereas all diagnostic methods had high specificity. There was perfect agreement in inter-examiner reliability for all assessment methods (Kappa 0.858–0.992). Conclusions: Both fluorescence-based devices and the visual-tactile method are reliable for detecting secondary caries around resin composite restorations. DIAGNOdent pen is accurate in enamel and dentin, while VistaCam and the visual-tactile method can detect secondary caries in enamel only. Clinical Relevance: Fluorescence-based devices could be used as a valuable aid to supplement or as a second opinion after the visual-tactile method. Trial registration: The study was listed on www.clinicaltrials.gov with registration number (NCT04426604) on 11/06/2020.Keywords
This publication has 49 references indexed in Scilit:
- A new universal simplified adhesive: 36-Month randomized double-blind clinical trialJournal of Dentistry, 2015
- Randomized clinical trial of four adhesion strategies in cervical lesions: 12-month results2015
- In vitro performance of QLF system and conventional methods for detection of occlusal caries around tooth‐colored restorations in primary molarsInternational Journal of Paediatric Dentistry, 2015
- Caries diagnosis using light fluorescence devices: VistaProof and DIAGNOdentOdontology, 2013
- Caries detection under composite restorations by laser fluorescence and digital radiographyClinical Oral Investigations, 2012
- The performance of conventional and fluorescence-based methods for occlusal caries detectionThe Journal of the American Dental Association, 2012
- Performance of laser fluorescence devices and visual examination for the detection of occlusal caries in permanent molarsJournal of Biomedical Optics, 2012
- In Vitro Detection of Secondary Caries Associated with Composite Restorations on Approximal Surfaces Using Laser FluorescenceOperative Dentistry, 2010
- Clinical performance of DIAGNOdent in the detection of secondary carious lesionsActa Odontologica Scandinavica, 2005
- Secondary caries detection by DIAGNOdent and radiography: a comparative in vitro studyActa Odontologica Scandinavica, 2004