Comparison of 4 CBCT-IGRT geometric QA phantoms within a single institutional workflow
- 21 August 2026
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 99 (1187) , 2261-2265
Abstract
Objectives: To compare 4 commercial cone-beam CT (CBCT) geometric quality assurance phantoms within a common workflow, focusing on positioning/repositioning accuracy and agreement between the laser-defined setup reference and the CBCT imaging isocenter. Methods: MIMI, QUASAR Penta-Guide, ISOCube, and MZ CUBE were evaluated monthly for 5 months on a Versa HD system with XVI CBCT and a HexaPOD couch. Two observers independently completed each measurement. Three-dimensional post-correction translational residuals (Δdpos/rep) and laser-defined setup reference-CBCT imaging isocenter distances (Δdiso) were calculated. Phantom comparisons used observer-averaged values from the 5 monthly sessions. Rotational correction and CBCT image quality were assessed descriptively with MZ CUBE. Results: Mean Δdpos/rep values ranged from 0.19 to 0.31 mm, with mean absolute observer differences of 0.05-0.12 mm. No overall difference was found among the 4 phantoms (P = .547). Mean Δdiso values ranged from 0.68 to 0.87 mm, and all measurements were below 2 mm. An overall difference was observed for Δdiso (P = .019), although no pairwise comparisons were performed. With MZ CUBE, mean post-correction translational and rotational residuals were 0.24 mm and 0.37°, respectively, and image-quality measures showed no consistent temporal trend. Conclusions: The 4 phantoms showed similarly small post-correction translational residuals. In our setting, MZ CUBE additionally supported rotational and image-quality checks. Advances in knowledge: Direct comparison within a common workflow showed similar translational performance, while the scope of additional testing depended on the phantom configuration and locally available accessories.Keywords
This publication has 13 references indexed in Scilit:
- Technical evaluation of a multipurpose phantom for selected CBCT QA endpoints across linac platformsPhysica Medica, 2026
- Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapyJournal of Applied Clinical Medical Physics, 2025
- AAPM Medical Physics Practice Guideline 8.b: Linear accelerator performance testsJournal of Applied Clinical Medical Physics, 2023
- AAPM Task Group 198 Report: An implementation guide for TG 142 quality assurance of medical acceleratorsMedical Physics, 2021
- Independent 6D quality assurance of stereotactic radiotherapy repositioning on linacsCancer/Radiothérapie, 2020
- Quality control in cone-beam computed tomography (CBCT) EFOMP-ESTRO-IAEA protocol (summary report)Physica Medica, 2017
- Quality assurance for image‐guided radiation therapy utilizing CT‐based technologies: A report of the AAPM TG‐179Medical Physics, 2012
- Isocenter verification for linac‐based stereotactic radiation therapy: review of principles and techniquesJournal of Applied Clinical Medical Physics, 2011
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Measurement of cone beam CT coincidence with megavoltage isocentre and image sharpness using the QUASAR™ Penta-Guide phantomPhysics in Medicine & Biology, 2008