Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy
Open Access
- 21 August 2025
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 26 (9) , e70239
Abstract
Purpose: This study presents MZ CUBE, a newly developed multipurpose phantom designed to assess both image quality and geometrical accuracy in image‐guided radiotherapy. Methods: The phantom is a 16 × 16 × 16 cm 3 cube made of acrylonitrile butadiene styrene and designed to evaluate HU constancy, HU uniformity, spatial integrity, spatial resolution, low contrast resolution, and geometrical accuracy. Two CT simulators (Big Bore RT and Discovery RT Gen3) and two cone‐beam CT (CBCT) systems (XVI and OBI) were used for verification alongside commercial phantoms (ACR 464 and CatPhan 504). Results: HU constancy differences were within 3 HU, and uniformity was within 1 HU for CT and 2 HU for CBCT, based on repeated MZ CUBE measurements. Spatial resolution with the MZ CUBE was 7 lp/cm (Big Bore RT), 8 lp/cm (Discovery RT Gen3), 7 lp/cm (XVI), and 8 lp/cm (OBI). The minimum discernible disk diameters were 7, 5, 6, and 6 mm, respectively. Geometrical accuracy was confirmed with translational offsets below 1 mm and rotational deviations below 0.2° for both CBCT systems. Conclusions: MZ CUBE enables comprehensive image quality and geometric accuracy evaluation in a single scan, offering an efficient and cost‐effective solution for radiotherapy QA.Keywords
Funding Information
- National Research Foundation of Korea
- Ministry of SMEs and Startups
This publication has 15 references indexed in Scilit:
- AAPM Task Group 198 Report: An implementation guide for TG 142 quality assurance of medical acceleratorsMedical Physics, 2021
- A phantom for testing Cone Beam CTsPhysical and Engineering Sciences in Medicine, 2020
- CBCT image quality QA: Establishing a quantitative programJournal of Applied Clinical Medical Physics, 2020
- Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233Medical Physics, 2019
- Quality control in cone-beam computed tomography (CBCT) EFOMP-ESTRO-IAEA protocol (summary report)Physica Medica, 2017
- Change in Image Quality According to the 3D Locations of a CBCT PhantomPLOS ONE, 2016
- Quality assurance for image‐guided radiation therapy utilizing CT‐based technologies: A report of the AAPM TG‐179Medical Physics, 2012
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Measurement of cone beam CT coincidence with megavoltage isocentre and image sharpness using the QUASAR™ Penta-Guide phantomPhysics in Medicine & Biology, 2008
- The phantom portion of the American College of Radiology (ACR) Computed Tomography (CT) accreditation program: Practical tips, artifact examples, and pitfalls to avoidMedical Physics, 2004