Right ventricular outflow tract septal pacing versus apical pacing: A prospective, randomized, single-blind 5-years follow-up study of ventricular lead performance and safety
- 16 December 2015
- journal article
- research article
- Published by Springer Nature in Current Medical Science
- Vol. 35 (6) , 858-861
Abstract
Lead placement for ventricular pacing variably impacts the physiological benefit of the patient. This study evaluated the ventricular lead performance and safety of right ventricular outflow tract septal pacing in patients with bradyarrhythmia in South China over 60-month follow-up. Totally, 192 patients (108 males, and 84 females, 63±21 years old) with bradyarrhythmia were randomly divided into two groups. The right ventricular outflow tract septum (RVOTs) group had lead placement near the septum (n=97), while the right ventricular apex (RVA) group had a traditional apical placement (n=95). RV septal lead positioning was achieved with a specialized stylet and confirmed using fluoroscopic projection. All patients were followed up for 60 months. Follow-up assessment included stimulation threshold, R-wave sensing, lead impedance and lead complications. The time of electrode implantation in both the ROVTs and RVA groups were significantly different (4.29±0.61 vs. 2.16±0.22 min; P=0.009). No differences were identified in threshold, impedance or R-wave sensing between the two groups at 1st, 12th, 36th and 60th month during the follow-up period. No occurrence of electrode displacement, increased pacing threshold or inadequate sensing was found. The long-term active fixation ventricular electrode performance in RVOTs group was similar to that in RVA group. RVOTs pacing near the septum using active fixation electrodes may provide stability during long-term follow-up period.Keywords
This publication has 16 references indexed in Scilit:
- Medium‐Term Effects of Septal and Apical Pacing in Pacemaker‐Dependent Patients: A Double‐Blind Prospective Randomized StudyPacing and Clinical Electrophysiology, 2013
- Pacing of the interventricular septum versus the right ventricular apex: A prospective, randomized studyEuropean Journal of Internal Medicine, 2012
- The Road to Right Ventricular Septal Pacing: Techniques and ToolsPacing and Clinical Electrophysiology, 2010
- Right Ventricular Septal Pacing: The Success of Stylet-Driven Active-Fixation LeadsPacing and Clinical Electrophysiology, 2009
- Selective site right ventricular pacingHeart, 2009
- The Effects of Right Ventricular Apical Pacing on Ventricular Function and DyssynchronyJACC, 2009
- Long-Term Impact of Right Ventricular Septal Versus Apical Pacing on Left Ventricular Synchrony and Function in Patients With Second- or Third-Degree Heart BlockThe American Journal of Cardiology, 2009
- Right Ventricular Outflow Tract Septal Pacing: Long‐Term Follow‐Up of Ventricular Lead PerformancePacing and Clinical Electrophysiology, 2009
- Long‐Term Performance of Active‐Fixation Pacing Leads: A Prospective StudyPacing and Clinical Electrophysiology, 2006
- Dual-Chamber Pacing or Ventricular Backup Pacing in Patients With an Implantable DefibrillatorJAMA, 2002