Single‐institution report of setup margins of voluntary deep‐inspiration breath‐hold (DIBH) whole breast radiotherapy implemented with real‐time surface imaging
Open Access
- 22 June 2018
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 19 (4) , 205-213
Abstract
Purpose: We calculated setup margins for whole breast radiotherapy during voluntary deep‐inspiration breath‐hold (vDIBH) using real‐time surface imaging (SI).Methods and Materials: Patients (n = 58) with a 27‐to‐31 split between right‐ and left‐sided cancers were analyzed. Treatment beams were gated using AlignRT by registering the whole breast region‐of‐interest to the surface generated from the simulation CT scan. AlignRT recorded (three‐dimensional) 3D displacements and the beam‐on‐state every 0.3 s. Means and standard deviations of the displacements during vDIBH for each fraction were used to calculate setup margins. Intra‐DIBH stability and the intrafraction reproducibility were estimated from the medians of the 5th to 95th percentile range of the translations in each breath‐hold and fraction, respectively.Results: A total of 7269 breath‐holds were detected over 1305 fractions in which a median dose of 200 cGy was delivered. Each fraction was monitored for 5.95 ± 2.44 min. Calculated setup margins were 4.8 mm (A/P), 4.9 mm (S/I), and 6.4 mm (L/R). The intra‐DIBH stability and the intrafraction reproducibility were ≤0.7 mm and ≤2.2 mm, respectively. The isotropic margin according to SI (9.2 mm) was comparable to other institutions’ calculations that relied on x‐ray imaging and/or spirometry for patients with left‐sided cancer (9.8–11.0 mm). Likewise, intra‐DIBH variability and intrafraction reproducibility of breast surface measured with SI agreed with spirometry‐based positioning to within 1.2 and 0.36 mm, respectively.Conclusions: We demonstrated that intra‐DIBH variability, intrafraction reproducibility, and setup margins are similar to those reported by peer studies who utilized spirometry‐based positioning.Keywords
Funding Information
- University of Chicago (P30 CA014599)
- National Institute on Aging (4T35AG029795‐10)
This publication has 32 references indexed in Scilit:
- Risk of Ischemic Heart Disease in Women after Radiotherapy for Breast CancerThe New England journal of medicine, 2013
- Six-year Experience Routinely Using Moderate Deep Inspiration Breath-hold for the Reduction of Cardiac Dose in Left-sided Breast Irradiation for Patients With Early-stage or Locally Advanced Breast CancerAmerican Journal of Clinical Oncology, 2013
- Accuracy Evaluation of a 3-Dimensional Surface Imaging System for Guidance in Deep-Inspiration Breath-Hold Radiation TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2013
- Radiation-related mortality from heart disease and lung cancer more than 20 years after radiotherapy for breast cancerBritish Journal of Cancer, 2012
- Image-Guided Radiotherapy for Left-Sided Breast Cancer Patients: Geometrical Uncertainty of the HeartInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Image-Guided Radiotherapy: Has It Influenced Patient Outcomes?Seminars in Radiation Oncology, 2012
- Comparison of Target Registration Errors for Multiple Image-Guided Techniques in Accelerated Partial Breast IrradiationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Image Guidance in External Beam Accelerated Partial Breast Irradiation: Comparison of Surrogates for the Lumpectomy CavityInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Errors and margins in radiotherapySeminars in Radiation Oncology, 2004
- Twenty-Year Follow-up of a Randomized Trial Comparing Total Mastectomy, Lumpectomy, and Lumpectomy plus Irradiation for the Treatment of Invasive Breast CancerThe New England journal of medicine, 2002