Direct and fast measurement of CT beam filter profiles with simultaneous geometrical calibration
Authors:
- 25 November 2016
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 44 (1) , 57-70
Abstract
Purpose To accurately measure the beam filter profiles from a variety of CT scanner models and to provide reference data for Monte Carlo simulations of CT scanners. Methods This study proposed a new method to measure CT beam filter profiles using a linear‐array x‐ray detector (X‐Scan 0.8f3‐512; Detection Technology Inc., Espoo, Finland) under gantry rotation mode. A robust geometrical calibration approach was developed to determine key geometrical parameters by considering the x‐ray focal spot location relative to the linear‐array detector and the gantry's angular increment at each acquisition point. CT beam intensity profiles were synthesized from continuously measured data during a 10° gantry rotation range with calibrated detector response and system geometry information. Relative transmission profiles of nineteen sets of beam filters were then derived for nine different CT scanner models from three different manufacturers. Equivalent aluminum thickness profiles of these beam filters were determined by analytical calculation using the Spektr Matlab software package to match the measured transmission profiles. Three experiments were performed to validate the accuracy of the geometrical calibration, detector response modeling, and the derived equivalent aluminum thickness profiles. Results The beam intensity profiles measured from gantry rotation mode showed very good agreement with those measured with gantry stationary mode, with a maximal difference of 3%. The equivalent aluminum thickness determined by this proposed method agreed well with what was measured by an ion chamber, with a mean difference of 0.4%. The determined HVL profiles matched well with data from a previous study (max difference of 4.7%). Conclusions An accurate and robust method to directly measure profiles from a broad list of beam filters and CT scanner models was developed, implemented, and validated. Useful reference data was provided for future research on CT system modeling.Keywords
Funding Information
- National Institute of Biomedical Imaging and Bioengineering (R01EB015478)
This publication has 21 references indexed in Scilit:
- Scatter radiation intensities around a clinical digital breast tomosynthesis unit and the impact on radiation shielding considerationsMedical Physics, 2016
- A new technique to characterize CT scanner bow‐tie filter attenuation and applications in human cadaver dosimetry simulationsMedical Physics, 2015
- A rapid noninvasive characterization of CT x-ray sourcesMedical Physics, 2015
- Technical Note: Measurement of bow tie profiles in CT scanners using radiochromic filmMedical Physics, 2015
- Patient-specific Radiation Dose and Cancer Risk for Pediatric Chest CTRadiology, 2011
- Experimental validation of a method characterizing bow tie filters in CT scanners using a real‐time dose probeMedical Physics, 2011
- A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT Monte Carlo dosimetry simulationsMedical Physics, 2009
- Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantomsMedical Physics, 2009
- Monte Carlo evaluation of in infinitely long cylinders of water, polyethylene and PMMA with diameters from toMedical Physics, 2008
- Monte Carlo simulations in CT for the study of the surface air kerma and energy imparted to phantoms of varying size and positionPhysics in Medicine & Biology, 2004