Comparison of corneal astigmatism measurements of 2 optical biometer models for toric intraocular lens selection
- 1 February 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 41 (2) , 364-371
Abstract
To compare corneal astigmatism measurements obtained using a new optical biometer (AL-Scan) and a standard biometer based on partial coherence interferometry (IOLMaster 500) for toric intraocular lens (IOL) selection.Siriraj Hospital, Mahidol University, Bangkok, Thailand.Prospective comparative study.Keratometry (K) readings were taken with the 2 biometers in eyes of cataract patients. The corneal astigmatism readings from the 2 devices were compared. The cylinder measurements were transformed into J0 and J45 vector components for analysis, and the devices' values were compared using the paired t test. Correlation coefficients were calculated. The Bland-Altman method was used to evaluate the agreement between the sets of measurements.One hundred thirty-seven eyes (81 patients) were enrolled. The magnitude of astigmatism between the new biometer 2.4 mm zone and the standard biometer was similar (P = .19). There was a statistically significant difference between values from the new biometer 3.3 mm zone and the standard biometer (P < .05). There was good correlation in the overall magnitude of astigmatism between devices. There was a trend of low correlation toward the flat cornea (K <43.0 diopters [D]), low corneal cylinder (<1.0 D), and with-the-rule astigmatism. The IOL toricity calculated from both devices was similar.Corneal astigmatism measurements with the 2 biometers were similar in magnitude, especially in 2.4 mm zone of the new biometer. There was no difference in the toric IOL selections between the devices.No author has a financial or proprietary interest in any material or method mentioned.Keywords
This publication has 23 references indexed in Scilit:
- Agreement Between Pentacam and IOLMaster in Patients Undergoing Toric IOL ImplantationJournal of Refractive Surgery, 2013
- Comparison of corneal astigmatism and axis location in cataract patients measured by total corneal power, automated keratometry, and simulated keratometryJournal of Cataract & Refractive Surgery, 2012
- Characteristics of Astigmatism as a Function of Age in a Hong Kong Clinical PopulationOptometry and Vision Science, 2012
- Comparison of corneal power, corneal astigmatism, and axis location in normal eyes obtained from an autokeratometer and a corneal topographerJournal of Cataract & Refractive Surgery, 2012
- Prevalence of corneal astigmatism in patients having routine cataract surgery at a teaching hospital in the United KingdomJournal of Cataract & Refractive Surgery, 2011
- Differences in corneal astigmatism between partial coherence interferometry biometry and automated keratometry and relation to topographic patternJournal of Cataract & Refractive Surgery, 2011
- Effects of preoperative corneal astigmatism orientation on results with a low-cylinder-power toric intraocular lensJournal of Cataract & Refractive Surgery, 2011
- Comparison of Corneal Powers Obtained from 4 Different DevicesAmerican Journal of Ophthalmology, 2009
- Meridional analysis for calculating the expected spherocylindrical refraction in eyes with toric intraocular lensesJournal of Cataract & Refractive Surgery, 2007
- Refractive Errors Induced by Displacement of Intraocular Lenses within the Pseudophakic EyeOptometry and Vision Science, 1989