Space-Specific Mixing Excitation for High-SNR Spatial Encoding in Magnetic Particle Imaging
- 13 May 2024
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 71 (10) , 2889-2899
Abstract
Objective: Magnetic Particle Imaging (MPI) is a radiation-free tracer-based imaging technology that visualizes the spatial distribution of superparamagnetic iron oxide nanoparticles. Conventional spatial encoding methods in MPI rely on a gradient magnetic field with a constant gradient strength to generate a field-free point or line for spatial scanning. However, increasing the gradient strength can enhance theoretical spatial resolution but also lead to a decrease in the Signal-to-Noise Ratio (SNR) and sensitivity of the imaging system. This poses a technical challenge in balancing spatial resolution and sensitivity, necessitating intricate hardware design. Methods: To address this, we present a Space-Specific Mixing Excitation (SSME) technique for achieving high-SNR spatial encoding in MPI. By utilizing a dual-frequency excitation magnetic field with a non-homogeneous field strength, magnetic particles at each position generate unique intermodulation responses. By performing multi-channel acquisitions across the entire field of view, high SNR MPI signals can be acquired. When combined with reconstruction techniques based on system matrix, multi-dimensional SSME-MPI can be achieved. Results: The effectiveness of the proposed method was validated through phantom and in vivo imaging experiments. The results demonstrate significant improvements in sensitivity (3.6-fold improvement) and spatial resolution (better than 1 mm) without any hardware modifications. Conclusion: These findings demonstrate the capability of SSME to enhance both the spatial resolution and sensitivity of MPI. Significance: This method provides a solution to the ongoing challenge of balancing spatial resolution and sensitivity in MPI, potentially facilitating the implementation of MPI in a wider range of medical applications.Keywords
Funding Information
- National Natural Science Foundation of China (62027901, 81930053)
- National Key Research and Development Program of China (2023YFC3402800)
- Beijing Natural Science Foundation (L232097)
- Fundamental Research Funds for the Central Universities (YWF23Q1072, YWF23Q1074)
- CAST (2023QNRC001)
- Key-Area Research and Development Program of Guangdong Province (2021B0101420005)
This publication has 38 references indexed in Scilit:
- Deep Penetrating and Sensitive Targeted Magnetic Particle Imaging and Photothermal Therapy of Early‐Stage Glioblastoma Based on a Biomimetic NanoplatformAdvanced Science, 2023
- Sensitive and specific detection of breast cancer lymph node metastasis through dual-modality magnetic particle imaging and fluorescence molecular imaging: a preclinical evaluationEuropean Journal of Nuclear Medicine and Molecular Imaging, 2022
- Monitoring Intracranial Cerebral Hemorrhage Using Multicontrast Real-Time Magnetic Particle ImagingACS Nano, 2020
- Magnetic Particle Imaging: Current Applications in Biomedical ResearchJournal of Magnetic Resonance Imaging, 2019
- Liposomal nanohybrid cerasomes targeted to PD-L1 enable dual-modality imaging and improve antitumor treatmentsCancer Letters, 2018
- Fast multiresolution data acquisition for magnetic particle imaging using adaptive feature detectionMedical Physics, 2017
- Molecular Imaging of Inflammation/Infection: Nuclear Medicine and Optical Imaging Agents and MethodsChemical Reviews, 2010
- Optimization of nanoparticle core size for magnetic particle imagingJournal of Magnetism and Magnetic Materials, 2009
- Magnetic particle detection by frequency mixing for immunoassay applicationsJournal of Magnetism and Magnetic Materials, 2007