Exchangeable Miniaturized Mass Spectrometer Chip Based on Silicon Structures
Open Access
- 30 November 2018
- conference paper
- Published by MDPI AG in Proceedings
- Vol. 2 (13) , 1071
Abstract
This paper reports a design of an exchangeable miniaturized mass spectrometry chip using spring-loaded pins and O-rings for electrical and fluidic connections. This planar microelectromechanical system (MEMS)-chip works with 300 µm high silicon structures between two borosilicate glasses and has a size of 13 mm × 7 mm. Because of its small size a small vacuum pump is sufficient, and it is suitable for mobile measurements and portable applications. Because of exchangeability of MEMS-chips with fluidic and electrical high voltage connections a direct comparison between chips is possible.Keywords
This publication has 2 references indexed in Scilit:
- Lab-on-a-chip devices: How to close and plug the lab?Microelectronic Engineering, 2015
- Mass spectra measured by a fully integrated MEMS mass spectrometerInternational Journal of Mass Spectrometry, 2007