Impact of peri-implant bone resorption, prosthetic materials, and crown to implant ratio on the stress distribution of short implants: a finite element analysis
- 17 March 2021
- journal article
- research article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 59 (4) , 813-824
Abstract
The purpose of this study was to determine the effects of prosthetic materials and crown/implant (C/I) ratio on short implants with a marginal bone resorption. Three-dimensional finite element analysis was used to simulate stress distribution under static loading in non-resorption and resorption scenarios (3-mm vertical bone loss) in implants restored with single crowns and C/I ratios of 1:1, 1.5:1, and 2:1 were evaluated. Different crown materials were used: porcelain-fused to metal, porcelain-fused to zirconia, monolithic zirconia, and zirconia-based crown veneered with indirect composite resin. The C/I ratio, the peri-implant bone resorption, and the loading conditions were the key factors affecting the generated stress in short implants. In non-resorption models, von Mises stress ranged between 50 and 105 MPa whereas in resorption models, the values ranged from 168 to 322 MPa, both increasing with the higher C/I ratio under oblique forces. Under axial loading, the C/I ratio did not influence the stress values as the presence of resorption was the only parameter increasing, 57 MPa for the non-resorption models and 101 MPa for the resorption models, respectively. Preference of a prosthetic material was ineffective on the distribution of stress in the bone and implant structure under static loading in any models. The peri-implant bone resorption and a higher C/I ratio in short implants increase the stress values under both axial and oblique forces, whereas the crown material does not influence stress distribution in the surrounding bone and implant structure. Graphical abstractKeywords
Funding Information
- Altınbaş University (PB2017/2)
This publication has 65 references indexed in Scilit:
- A three-dimensional finite element analysis of the stress distribution on morse taper implants surfaceJournal of Prosthodontic Research, 2013
- Increased Crown-To-Implant Ratio May Not Be a Risk Factor for Dental Implant Failure under Appropriate Plaque ControlPLOS ONE, 2013
- Comparisons of maximum deformation and failure forces at the implant–abutment interface of titanium implants between titanium-alloy and zirconia abutments with two levels of marginal bone lossBioMedical Engineering OnLine, 2013
- Short implants: A systematic reviewJournal of Indian Society of Periodontology, 2012
- 3D Finite Element Analysis of Changes in Stress Levels and Distributions for an Osseointegrated Implant After Vertical Bone LossImplant Dentistry, 2011
- Crown-to-Implant Ratios of Short-Length ImplantsJournal of Oral Implantology, 2010
- Implant Platform SwitchingJournal of Craniofacial Surgery, 2010
- Influence of forces on peri‐implant boneClinical Oral Implants Research, 2006
- A Targeted Review of Study Outcomes With Short (≤7 mm) Endosseous Dental Implants Placed in Partially Edentulous PatientsJournal of Periodontology, 2004
- Finite element analysis of non‐axial versus axial loading of oral implants in the mandible of the dogJournal of Oral Rehabilitation, 1998