Effect of remote monitoring on clinical outcomes in European heart failure patients with an implantable cardioverter-defibrillator: secondary results of the REMOTE-CIED randomized trial
Authors:
- Cheyenne S L ChiuCCCheyenne S L Chiu
- Mathias MeineMMMathias Meine
- for the REMOTE-CIED Trial InvestigatorsFIfor the REMOTE-CIED Trial Investigators
Open Access
- 19 August 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 24 (2) , 256-267
Abstract
Remote patient monitoring (RPM) systems offer a promising alternative to conventional In-Clinic check-ups, hereby reducing unnecessary clinic visits. Especially with the rise of the COVID-19 pandemic, this reduction is of paramount importance. Regarding the association between RPM and clinical outcomes, findings of previous studies have been inconsistent. The aim of this study is to elucidate the effect of partly substituting In-Clinic visits by RPM on clinical outcomes in implantable cardioverter-defibrillator (ICD) patients. The study included 595 heart failure patients (LVEF ≤35%; NYHA Class II/III) implanted with an ICD compatible with the Boston Scientific LATITUDE™ system. Participants were randomized to RPM plus an annual In-Clinic visit or 3–6 months In-Clinic check-ups alone. The investigated endpoints after 2 years of follow-up included a composite of all-cause mortality and cardiac hospitalization, mortality and cardiac hospitalization as independent endpoints and ICD therapy. The incidence of mortality and hospitalization did not differ significantly as independent, nor as composite endpoint between the RPM and In-Clinic group (all Ps P = 0.04). Exploratory subgroup analyses indicated that the effect of RPM differs between patients with specific characteristics, i.e. ≥60 years and permanent atrial fibrillation (all Ps < 0.05). RPM is non-inferior to conventional In-Clinic visits regarding clinical outcomes. Routine In-Clinic follow-up may partly be substituted by RPM without jeopardizing safety and efficiency, and thus reducing unnecessary In-Clinic visits. NCT01691586.Keywords
Funding Information
- Boston Scientific
This publication has 24 references indexed in Scilit:
- Effect of Telemonitoring of Cardiac Implantable Electronic Devices on Healthcare Utilization: A Meta-Analysis of Randomized Controlled Trials in Patients with Heart FailureEuropean Journal of Heart Failure, 2016
- HRS Expert Consensus Statement on remote interrogation and monitoring for cardiovascular implantable electronic devicesHeart Rhythm, 2015
- Remote Monitoring of Implantable Cardioverter-DefibrillatorsJACC, 2015
- Patient perspective on remote monitoring of cardiovascular implantable electronic devices: rationale and design of the REMOTE-CIED studyNetherlands Heart Journal, 2014
- 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: The Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA)EP Europace, 2013
- Effect of Long-Detection Interval vs Standard-Detection Interval for Implantable Cardioverter-Defibrillators on Antitachycardia Pacing and Shock DeliveryJAMA, 2013
- Implementation of guidelines for implantable cardioverter-defibrillator therapy in clinical practice: Which patients do benefit?Netherlands Heart Journal, 2013
- Reduction in Inappropriate Therapy and Mortality through ICD ProgrammingNew England Journal of Medicine, 2012
- A randomized study of remote follow-up of implantable cardioverter defibrillators: safety and efficacy report of the ECOST trialEuropean Heart Journal, 2012
- ISHNE/EHRA expert consensus on remote monitoring of cardiovascular implantable electronic devices (CIEDs)EP Europace, 2012