Design of a head phantom produced on a 3D rapid prototyping printer and comparison with a RANDO and 3M lucite head phantom in eye dosimetry applications
- 21 March 2017
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (8) , 3158-3174
Abstract
An anthropomorphic head phantom including eye inserts allowing placement of TLDs 3 mm below the cornea has been produced on a 3D printer using a photo-cured acrylic resin to best allow tissue equivalence. Thus H p (3) can be determined in radiological and interventional photon radiation fields. Eye doses and doses to the forehead have been compared to an Alderson RANDO head and a 3M Lucite skull phantom in terms of surface dose per incident air kerma for frontal irradiation since the commercial phantoms do not allow placement of TLDs 3 mm below the corneal surface. A comparison of dose reduction factors (DRFs) of a common lead glasses model has also been performed. Eye dose per incident air kerma were comparable between all three phantoms (printed phantom: 1.40, standard error (SE) 0.04; RANDO: 1.36, SE 0.03; 3M: 1.37, SE 0.03). Doses to the forehead were identical to eye surface doses for the printed phantom and the RANDO head (ratio 1.00 SE 0.04, and 0.99 SE 0.03, respectively). In the 3M Lucite skull phantom dose on the forehead was 15% lower than dose to the eyes attributable to phantom properties. DRF of a sport frame style leaded glasses model with 0.75 mm lead equivalence measured were 6.8 SE 0.5, 9.3 SE 0.4 and 10.5 SE 0.5 for the RANDO head, the printed phantom, and the 3M Lucite head phantom, respectively, for frontal irradiation. A comparison of doses measured in 3 mm depth and on the surface of the eyes in the printed phantom revealed no difference larger than standard errors from TLD dosimetry. 3D printing offers an interesting opportunity for phantom design with increasing potential as printers allowing combinations of tissue substitutes will become available. Variations between phantoms may provide a useful indication of uncertainty budgets when using phantom measurements to estimate individual personnel doses.Keywords
This publication has 34 references indexed in Scilit:
- Multi-resolution voxel phantom modeling: a high-resolution eye model for computational dosimetryPhysics in Medicine & Biology, 2014
- Public Health England survey of eye lens doses in the UK medical sectorJournal of Radiological Protection, 2013
- Report of Task Group on the implications of the implementation of the ICRP recommendations for a revised dose limit to the lens of the eyeJournal of Radiological Protection, 2013
- Eye lens dosimetry in interventional cardiology: results of staff dose measurements and link to patient dose levelsRadiation Protection Dosimetry, 2012
- On the operational quantityHp(3) for eye lens dosimetryJournal of Radiological Protection, 2012
- Backscatter factors and mass energy-absorption coefficient ratios for diagnostic radiology dosimetryPhysics in Medicine & Biology, 2011
- Dose conversion coefficients for photon exposure of the human eye lensPhysics in Medicine & Biology, 2010
- Monitoring the eye lens: which dose quantity is adequate?Physics in Medicine & Biology, 2010
- Dose conversion coefficients for electron exposure of the human eye lensPhysics in Medicine & Biology, 2009
- Measurement of backscattered x-ray spectra at the water surface in the energy range 60 kV to 120 kVPhysics in Medicine & Biology, 2002