Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients
- 16 November 2012
- journal article
- research article
- Published by Springer Nature in International Ophthalmology
- Vol. 33 (3) , 221-225
Abstract
To compare intraocular pressure (IOP) measurements obtained with Goldmann applanation (GAT) and dynamic contour tonometry (DCT) in a Mexican population. 40 glaucoma patients were included in this cross-sectional observational cohort study. IOP measurements were performed in the following order: DCT, ultrasonic pachymetry and GAT, with a 5-minute difference between each measurement, between 8 am and 2 pm. Only DCT measurements of good quality (Q ≤ 3) were accepted. GAT measurements were made three times with the same Goldmann tonometer, previously checked for calibration errors, and the mean was used for statistical purposes. The IOP (mean [standard deviation], 95 % confidence interval [CI]) measured with the Goldmann tonometer (13.2 [2.4], 12.4–14.0 mmHg) was significantly lower than that obtained with the DCT (18.4 [3.3], 17.0–19.2 mmHg), p r = 0.24, 95 % CI = 0.07–0.52, p = 0.133) and (r = 0.13, 95 % CI = –0.19 to 0.43, p = 0.412), respectively. The concordance correlation coefficient between GAT and DCT was r c = 0.3, 95 % CI = 0.17–0.41). DCT seems to overestimate the IOP as compared to GAT. Additionally, although there was a good correlation between the IOP measurements assessed with either GAT or DCT, the agreement was poor.Keywords
This publication has 26 references indexed in Scilit:
- Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practiceActa Ophthalmologica, 2009
- Effect of Central Corneal Thickness and Corneal Hysteresis on Tonometry as Measured by Dynamic Contour Tonometry, Ocular Response Analyzer, and Goldmann Tonometry in Glaucomatous EyesJournal of Glaucoma, 2008
- New ways to measure intraocular pressureCurrent Opinion in Ophthalmology, 2008
- Comparisons between Pascal dynamic contour tonometry, the TonoPen, and Goldmann applanation tonometry in patients with glaucomaActa Ophthalmologica Scandinavica, 2006
- Comparison of Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Glaucoma Patients and Healthy SubjectsAmerican Journal of Ophthalmology, 2006
- The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometryBritish Journal of Ophthalmology, 2005
- Dynamic Contour TonometryJournal of Glaucoma, 2005
- Dynamic Contour Tonometry for Intraocular Pressure MeasurementAmerican Journal of Ophthalmology, 2005
- Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometryEye, 2005
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986