Influence of magnetic field on a novel scintillation dosimeter in a 1.5 T MR‐linac
Open Access
- 27 November 2023
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (1) , e14180
Abstract
For commissioning and quality assurance for adaptive workflows on the MR‐linac, a dosimeter which can measure time‐resolved dose during MR image acquisition is desired. The Blue Physics model 10 scintillation dosimeter is potentially an ideal detector for such measurements. However, some detectors can be influenced by the magnetic field of the MR‐linac. To assess the calibration methods and magnetic field dependency of the Blue Physics scintillator in the 1.5 T MR‐linac. Several calibration methods were assessed for robustness. Detector characteristics and the influence of the calibration methods were assessed based on dose reproducibility, dose linearity, dose rate dependency, relative output factor (ROF), percentage depth dose profile, axial rotation and the radial detector orientation with respect to the magnetic field. The potential application of time‐resolved dynamic dose measurements during MRI acquisition was assessed. A variation of calibration factors was observed for different calibration methods. Dose reproducibility, dose linearity and dose rate stability were all found to be within tolerance and were not significantly affected by different calibration methods. Measurements with the detector showed good correspondence with reference chambers. The ROF and radial orientation dependence measurements were influenced by the calibration method used. Axial detector dependence was assessed and relative readout differences of up to 2.5% were observed. A maximum readout difference of 10.8% was obtained when rotating the detector with respect to the magnetic field. Importantly, measurements with and without MR image acquisition were consistent for both static and dynamic situations. The Blue Physics scintillation detector is suitable for relative dosimetry in the 1.5 T MR‐linac when measurements are within or close to calibration conditions.Keywords
Funding Information
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek (18495)
- European Metrology Programme for Innovation and Research (19NRM01)
This publication has 32 references indexed in Scilit:
- Dosimetry in clinical static magnetic fields using plastic scintillation detectorsRadiation Measurements, 2013
- Towards reference dosimetry for the MR-linac: magnetic field correction of the ionization chamber readingPhysics in Medicine & Biology, 2013
- Plastic scintillation dosimetry: comparison of three solutions for the Cerenkov challengePhysics in Medicine & Biology, 2011
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- MRI/linac integrationRadiotherapy and Oncology, 2008
- Experimental verification of magnetic field dose effects for the MRI-acceleratorPhysics in Medicine & Biology, 2007
- Plastic scintillation dosimetry and its application to radiotherapyRadiation Measurements, 2006
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Cerenkov light generated in optical fibres and other light pipes irradiated by electron beamsPhysics in Medicine & Biology, 1992