A Diode-Based Symmetric Charge Division Circuit With Grounding Path to Reduce Signal Crosstalk and Improve Detector Performance
- 13 September 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Radiation and Plasma Medical Sciences
- Vol. 6 (7) , 788-793
Abstract
Multiplexing circuits in pixelated PET detectors are being studied with the goal of minimizing the degradation of detector performance. The purpose of this study was to develop a diode-based symmetric charge division (SCD) circuit with reducing the crosstalk caused by the leakage currents of adjacent channels connected in rows and columns. Four different SCD circuits (diode-based SCD with grounding path, diode-based SCD, resistor-based SCD with grounding path, and resistor-based SCD) were constructed and their performance was evaluated and compared after being combined with a PET detector consisting of Lutetium Yttrium Orthosilicate crystals optically coupled with SiPM. The diode-based SCD circuit with grounding path achieved a crosstalk ratio, energy resolution, coincidence timing resolution (CTR), and mean peak to valley ratio of 0.02, 14.3%, 0.5 ns, and 27.9: 1, respectively; the diode-based SCD circuit achieved 0.03, 15.9%, 0.7 ns, and 19.4: 1, respectively; the resistor-based SCD circuit with grounding path achieved 0.04, 14.5%, 0.9 ns, and 19.4: 1, respectively; and the resistor-based SCD circuit achieved 0.05, 17.8%, 1.8 ns, and 17.7: 1, respectively. These results indicate that the proposed SCD has the potential to improve not only the crosstalk but also the energy linearity, energy resolution, and CTR characteristics in PET detectors.Keywords
Funding Information
- Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (2019R1I1A1A01041347)
- Korea Medical Device Development Fund grant funded by the Korea government (202011B11)
- Korea Medical Device Development Fund grant funded by the Korea government (202011A06)
This publication has 23 references indexed in Scilit:
- Development of a high-density position decoder circuit for GAPD-based PET with a large number of readout channelsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- Influence of detector pixel size, TOF resolution and DOI on image quality in MR-compatible whole-body PETPhysics in Medicine & Biology, 2013
- Investigation of analog charge multiplexing schemes for SiPM based PET block detectorsPhysics in Medicine & Biology, 2013
- Comparison of three resistor network division circuits for the readout of 4×4 pixel SiPM arraysNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- Simulation Study of Resistor Networks Applied to an Array of 256 SiPMsIEEE Transactions on Nuclear Science, 2013
- Free-running ADC- and FPGA-based signal processing method for brain PET using GAPD arraysNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2012
- Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR ScannerJournal of Nuclear Medicine, 2011
- A simple and improved digital timing method for positron emission tomographyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- A novel readout concept for multianode photomultiplier tubes with pad matrix anode layoutNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- Simple charge division readouts for imaging scintillator arrays using a multi-channel PMTIEEE Transactions on Nuclear Science, 1996