Enhancing the Usability of Patient Monitoring Devices in Intensive Care Units: Usability Engineering Processes for Early Warning System (EWS) Evaluation and Design
Open Access
- 6 May 2025
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 14 (9) , 3218
Abstract
Background/Objectives: This study aimed to enhance the usability of patient monitoring systems by integrating the Early Warning Score (EWS) function and improving user interface elements. The EWS function is expected to enable the early detection of acute deterioration and prompt medical intervention, while the optimized design supports rapid decision-making by nursing staff. Methods: Two formative usability evaluations were conducted to identify user requirements and improve the device design. A simulated usability test involved five ICU medical staff members, followed by a user preference survey with 72 ICU staff members in a real clinical setting. After incorporating feedback, a summative usability test with 23 ICU nurses was performed to evaluate the revised device. Results: Issues related to unfamiliar parameter terminology and alarm message positioning were identified, and the need for the EWS function was emphasized. The summative evaluation showed an increase in task success rate from 86% to 90% and a significant improvement in user satisfaction from 74.85 (SD: 0.88) to 89.55 (SD: 0.75) (p < 0.05). Conclusions: The integration of the EWS function and interface improvements significantly enhanced the usability of patient monitoring system. These advancements are expected to enable rapid detection of patient deterioration and support timely clinical decision-making by ICU staff.Keywords
Funding Information
- Korean government Korea Medical Device Development Fund (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health and Welfare, and the Ministry of Food and Drug Safety) (RS-2020-KD000030)
This publication has 22 references indexed in Scilit:
- Feasibility study of the use of a wearable vital sign patch in an intensive care unit settingJournal of Clinical Monitoring and Computing, 2024
- Usability and user experience of medical devices: An overview of the current state, analysis methodologies, and future challengesInternational Journal of Industrial Ergonomics, 2020
- Being there again – Presence in real and virtual environments and its relation to usability and user experience using a mobile navigation taskInternational Journal of Human-Computer Studies, 2017
- Exploring the performance of the National Early Warning Score (NEWS) in a European emergency departmentResuscitation, 2015
- The impact of the use of the Early Warning Score (EWS) on patient outcomes: A systematic reviewResuscitation, 2014
- Impact of critical care reconfiguration and track-and-trigger outreach team intervention on outcomes of haematology patients requiring intensive care admissionAnnals of Hematology, 2009
- Incidence, staff awareness and mortality of patients at risk on general wardsResuscitation, 2008
- Effect of introducing the Modified Early Warning score on clinical outcomes, cardio‐pulmonary arrests and intensive care utilisation in acute medical admissions*Anaesthesia, 2003
- Evaluation of product usability: development and validation of usability dimensions and design elements based on empirical modelsInternational Journal of Industrial Ergonomics, 2000